Back to Journals » Journal of Pain Research » Volume 16

Creating Realistic Definitions of Clinically Significant Radiographic Lead Migration – A Response to “Migration of Epidural Leads During Spinal Cord Stimulator Trials” [Letter]

Authors Mullins CF 

Received 8 January 2023

Accepted for publication 17 January 2023

Published 19 January 2023 Volume 2023:16 Pages 151—152

DOI https://doi.org/10.2147/JPR.S402488

Checked for plagiarism Yes

Editor who approved publication: Dr Andrea Tinnirello



Cormac F Mullins

Guy’s and St. Thomas’s NHS Foundation Trust, London, UK

Correspondence: Cormac F Mullins, Email [email protected]


View the original paper by Dr Jenkinson and colleagues

A Response to Letter has been published for this article.


Dear editor

I read with interest the retrospective study entitled “Migration of Epidural Leads During Spinal Cord Stimulator Trials” published in Journal of Pain Research.1

I would like to commend the authors for drawing attention to the underrecognised problem of lead migration during SCS trials. While lead migration has been reported as the most common complication of SCS and can result in failure to stimulate the intended area of the spinal cord with resultant loss of efficacy, lead migration during SCS trials presents a different set of challenges.2 Firstly, response to stimulation may not yet have been established during an SCS trial and therefore lead migration may result in a false-negative SCS trial. Secondly, with the advent of paraesthesia-independent waveforms, loss of paraesthesia is now a poor marker of lead migration. Thirdly, many clinicians do not perform radiography at the end of the trial in order to confirm that lead migration has not taken place.3 This is an important measure to facilitate subsequent surgical planning but also to out rule the presence of lead migration that can contribute to false negative and false-positive trials. If significant lead migration has taken place in the context of a negative trial, then consideration should be given to repeating the trial. Similarly, if there has been significant lead migration in the context of a positive trial, the clinician should consider the implications of this. It may be the case that the lead migrated immediately prior to the radiograph. If only one of two leads has migrated significantly, stimulation can still be delivered effectively on a single unmigrated lead. However, if the stimulation has been delivered on a lead that has migrated significantly, the contribution of a placebo effect resulting in a false-positive trial cannot be out ruled.

Defining what constitutes clinically significant lead migration is important in order to approach the problem correctly. The definition of clinically significant lead migration proposed by the authors of “greater than or equal to 50% of a full vertebral level” in at least one lead is problematic for a number of reasons.1 First of all, 94% of participants in the study met this criterion, suggesting that this is a clinically expected degree of lead migration. All of these leads had migrated caudally. In practice, most physicians expect that there will be some caudal migration of percutaneous leads during a SCS trial and site their slightly higher leads to anticipate this. Secondly, the authors outline that “50% of a vertebral level would typically correlate with roughly 1 centimetre of migration assuming an average vertebral body and disc height of 24 mm in the mid- and lower thoracic spine”. The distance covered by electrodes on SCS leads vary depending on the manufacturer; Nevro Octrode leads are 59 mm in length while Boston Scientific Linear 3–4 leads are 52 mm, Medtronic Vectris 3–4 leads are 52 mm and Abbott Quatrode 3–4 leads are 52 mm. If the midpoint of a 3–4 SCS lead (52 mm electrode coverage) was placed over the target area of stimulation (at 26 mm), lead migration of 10 mm would keep the target area easily inside the area of deliverable stimulation. Therefore, depending on the SCS lead employed, migration of at least 20 mm (or a full vertebral level) would be required in order to constitute clinically significant lead migration. Thirdly, as stated above, while one lead may migrate significantly during a trial, if two leads have been sited, stimulation can often be delivered on the remaining unmigrated lead, depending on the waveform and the configuration of leads.

I would suggest an alternative definition for clinically significant radiographic SCS lead migration of “lead migration of all leads outside the intended area of stimulation”. For ease of rapid clinical decision-making, this would typically correspond to migration of approximately half the distance covered by the electrodes of a standard length octopolar SCS lead (52–59 mm) which is at least a full vertebral level in the thoracic spine or two vertebral levels in the cervical spine.4,5 Investigating the incidence of clinically significant radiographic lead migration at the end SCS trials is an important scientific endeavour, especially as it may affect trial success, however, creating realistic definitions of lead migration is an essential step in this process.

Disclosure

The author reports no conflicts of interest in this communication.

References

1. Jenkinson RH, Wendahl A, Zhang Y, Sindt JE. Migration of epidural leads during spinal cord stimulator trials. J Pain Res. 2022;15:2999–3005. doi:10.2147/JPR.S378937

2. Kumar K, Buchser E, Linderoth B, Meglio M, Van Buyten JP. Avoiding complications from spinal cord stimulation: practical recommendations from an international panel of experts. Neuromodulation. 2007;10(1):24–33. doi:10.1111/j.1525-1403.2007.00084.x

3. Mekhail NA, Mathews M, Nageeb F, Guirguis M, Mekhail MN, Cheng J. Retrospective review of 707 cases of spinal cord stimulation: indications and complications. Pain Pract. 2011;11(2):148–153. doi:10.1111/j.1533-2500.2010.00407.x

4. Prameela M, Prabhu LV, Murlimanju B, Pai MM, Rai R, Kumar CG. Anatomical dimensions of the typical cervical vertebrae and their clinical implications. Eur J Anat. 2020;24(1):9–15.

5. Kunkel ME, Herkommer A, Reinehr M, Böckers TM, Wilke HJ. Morphometric analysis of the relationships between intervertebral disc and vertebral body heights: an anatomical and radiographic study of the human thoracic spine. J Anat. 2011;219(3):375–387. doi:10.1111/j.1469-7580.2011.01397.x

Creative Commons License © 2023 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.