Assessment of Trend, Indication, Complications, and Outcomes of Pacemaker Implantation in Adult Patients at Tertiary Hospital of Ethiopia: Retrospective Follow Up Study

Authors Markos S , Nasir M , Ahmed M , Abebe S, Amogne MA, Tesfaye D , Mekonnen TS , Getachew YG

Received 4 November 2023

Accepted for publication 10 January 2024

Published 11 January 2024 Volume 2024:17 Pages 93—103

DOI https://doi.org/10.2147/IJGM.S448135

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 2

Editor who approved publication: Prof. Dr. Yuriy Sirenko



Sura Markos,1 Mohammed Nasir,2 Muluken Ahmed,3 Sintayehu Abebe,4 Mebratu Ayele Amogne,4 Demu Tesfaye,4 Tigist Seleshi Mekonnen,4 Yitagesu Getachew Getachew5

1Internal Medicine Department, Hawassa University, Hawassa, Ethiopia; 2Pediatrics Department, Hawassa University, Hawassa, Ethiopia; 3Pediatrics Department, Arba Minch University, Arba Minch, Ethiopia; 4Internal Medicine Department, Addis Ababa University, Addis Ababa, Ethiopia; 5Internal Medicine Department, Yekatit 12 Teaching Hospital, Addis Ababa, Ethiopia

Correspondence: Sura Markos, Internal Medicine Department, Hawassa University, Hawassa, Ethiopia, Email [email protected]

Background: A pacemaker is a device implanted in the chest to help people with symptomatic bradycardia and life-threatening irregular rhythm. However, it is also associated with many complications. Therefore, this study evaluated pacemaker-related complications and factors associated with them, as there is little data on pacemaker-related complications in sub-Saharan Africa and Ethiopia.
Methods: The study was conducted on 118 patients over 18 years old who had pacemakers implanted between 2017 and 2022 at Tikur Anbessa Comprehensive Specialized Hospital in Addis Ababa, Ethiopia who were reviewed by the authors from September 2022 to December 2022. Sociodemographic factors, clinical characteristics, and complications data were extracted using a structured questionnaire by retrospective review of patient records. The chi-square test or Fisher’s exact test was performed to evaluate factors associated with complications.
Results: The median age of patients was 60.5 years (IQR = 15 years), with men accounting for 50.8% of patients. Hypertension was the most common comorbidity (64.2%). Symptomatic grade 3 AV block was the most common indication (78.8%) for pacemaker implantation. With a mean follow-up of 3.92 ± 1.94 years, 15.3% of patients had complications. Pneumothorax, pocket site infection, and lead dislodgement were the most common complications occurring in 2.54% of patients each. Patient age during surgery (p-value = 0.02), patient gender (p-value = 0.04), pacemaker implanting team (p-value = 0.01), and adherence to follow-up (p-value = 0.04) are related to pacemakers-related complications.
Conclusion: Pacemaker implantation is associated with many complications. Pneumothorax, pacemaker pocket infection, and lead dislodgement were the most common complications. Patient age at pacemaker implantation, patient gender, pacemaker implanting team, and follow-up compliance were factors associated with pacemaker-related complications. Skill development through specialized training and compliance counseling may improve outcomes for patients who have complications related to pacemaker implanting team and poor adherence to follow.

Keywords: pacemaker, pacemaker-related complications, associated factors

Introduction

A pacemaker is a device that is implanted in the chest to assist people with symptomatic bradycardia and life-threatening irregular rhythm.1 Due to the population’s increased life expectancy, its implantation rate has substantially increased globally.2 It not only contributes to certain people’s survival but also improves the quality of life for the majority of people, which explains its dramatic implantation increment.2 The most common indications for permanent pacemaker therapy worldwide include high-grade atrioventricular block (AVB) and sinus node dysfunction.1,2

Even though pacemakers are increasingly used globally to improve the quality of life and lengthen the lives of patients with life-threatening arrhythmia and symptomatic bradycardia, they are associated with several complications.2 Some examples of complications include lead failure, lead dislodgement, thoracic trauma, vascular injury, pocket hematoma, infection, pneumothorax, infection, thrombophlebitis, and death.3–7

Although understanding the magnitude of pacemaker-related complications is essential for initiatives to raise the standard of care, there are surprisingly few studies that assess their frequency in sub-Saharan nations, including none in Ethiopia. Therefore, the objective of this study is to evaluate complications related to pacemaker implantation and its associated factors in patients who underwent the procedure at Tikur Anbessa Comprehensive Hospital in Ethiopia from January 2017 to January 2022.

Methods

This study was conducted at Tikur Anbessa Comprehensive Specialized Hospital in Addis Ababa, Ethiopia who were reviewed by author between September 2022 and December 2022. All patients who met the criteria for inclusion were included. A follow-up study was employed with a retrospective chart review study design applied.

Eligibility Criteria

Inclusion criteria was all adult cardiac patients with permanent pacemakers older than 18 years old were treated at Tikur Anbessa Specialized Hospital is inclusion criteria. Patients with incomplete medical records or charts, those under the age of 18, those with ICDs, and those who had CRT were excluded.

Study Variables

Dependent variables were indication and complication of pace maker insertion, outcome, experience of the implanting team and adherence to follow up.

Independent variable were age and sex.

Operative Techniques and Procedures

The left Infraclavicular fossa with antepectoral plane was the default pocket for placing the generator with 1% Lidocaine local anesthetic, unless the patient had a history of pocket site infection in which case the right Infraclavicular fossa side would be chosen. The left subclavian vein was used for the insertion of the leads in 80 patients (67.8%), the left axillary vein in 35 patients (29.7%), the right subclavian vein in 2 patients, and the left cephalic vein in 1 patient. Prophylactic antibiotic with Ceftriaxone 1gm IV or Cefazolin 1gm IV was administered to all patients 1-hour before the procedure. In all cases active fixation was done. Anticoagulants were discontinued before the procedure without IV anticoagulant bridging. The pacemakers were checked as routine on the next day of implantation with cardiologist as part of postoperative follow up before discharge.

All patient had postoperative Chest X-ray before discharge. More than 85% of patients were discharged on the next day of implantation.

The Following Terms and Operational Definitions are Used

Early battery failure is the term used to describe a battery that fails before the manufacturer’s specified lifespan. In our study, a drop in left ventricular ejection fraction below 40% in a patient whose initial ejection fraction was higher than 40% was referred to as pacemaker-induced cardiomyopathy. The patient was labeled as adherent if he/she did not miss any clinic follow-up sessions; otherwise, he/she was labeled as non-adherent.

Data Collection Tool and Procedure

Data was gathered using a structured questionnaire. After reviewing numerous works of literature, the instrument was developed. Information for this study was gathered by reviewing patient’s medical records. Data collection and management were carried out by healthcare professionals who are knowledgeable and experienced. A two-day training was given to both the supervisors and the data collectors. The training put a lot of emphasis on the objectives of the study, the content of the tool, the technique for gathering data, ethical issues, and the responsibilities of data collectors.

Data Quality Control

To ensure the quality of the data, the focus was placed on well-designed data collection tools. Supervisors and data gatherers both got training. To ensure validity and consistency, the instrument was pre-tested on 5% of the sample outside the study area. The lead researcher oversaw and guided the entire data-gathering procedure.

Data Processing and Analysis

To confirm the accuracy of the data, it was manually reviewed. The data was coded, and then after data cleaning in the Epi-data program version 4.4.2.1, it was exported to SPSS for Windows, version 25 (SPSS, Chicago, IL, USA) for analysis.

The normality of continuous variables was examined using the Shapiro–Wilk test. The choice was made based on the findings of the normality test, and then the pertinent descriptive statistics for continuous variables were carried out. To display categorical variables, which were described using absolute frequency and percentages, tables and graphs were employed. Based on appropriateness, the chi-square test or Fisher exact test was applied to evaluate the association of the dependent variable with the independent variables. To determine statistical significance, a p-value of less than 0.05 was used.

Results

Sociodemographic Characteristics of the Patients

Out of 128 patients 118 were included in this study, with 10 patients excluded because of an incomplete chart. More than three-quarters of the 118 patients who were included in this study fell within the age range of 23–70 years, with a median age of 60.5 years (IQR: 52 years to 67 years). More than half of the patients were from Addis Ababa (Table 1).

Table 1 Shows the Sociodemographic and Clinical Characteristics of Patients Who Underwent Pacemaker Implantation in Tikur Anbessa Hospital from 2017 to 2022

The mean follow-up period of the study participants was 3.92±1.94 years. The majority of patients (53.4%) experienced light-headedness, syncope, and mild dyspnoea before pacemaker implantation, with hypertension being the most common comorbid condition (64.2%). Less than half of patients (28.8%) showed normal echocardiographic findings, and a significant percentage of patients (84.7%) had no history of heart failure. Only 6.8% of patients were using oral anticoagulants (warfarin or rivaroxaban), while 22.9% of patients were also taking enalapril, amlodipine, and aspirin. Symptomatic third-degree AV block was the most frequent indication (78.8%) for pacemaker implantation. A local team implanted more than half of the pacemakers, and more than three-quarters of them were done so for the first time. The majority (79.7% of cases) used dual chamber mode pacing, with more than half of the cases using a generator under the Medtronic brand. With a complication rate of 15.3%, more than half of the patients do not adhere to follow-up (Figure 1 and Table 1).

Figure 1 Number and type of complications of patients who underwent pacemaker implantation in Tikur Anbessa Hospital from 2017 to 2022.

Age of the patient, sex, symptoms before pacemaker implantation, type of operating team, and adherence to follow-up were the factors associated with pacemaker implantation complications (Table 2).

Table 2 Shows Factors Associated with Complication of Patients Who Underwent Pacemaker Implantation in Tikur Anbessa Hospital from 2017 to 2022

Discussion

This study evaluated the rate of pacemaker-related complications and its associated factors. Comparable complication rate to other studies were found, and multiple associated factors like patient age, patient sex, pacemaker implanting team, and follow-up adherence were also identified. We identified also the most common pacemaker-related complications as pneumothorax, lead dislodgement, and pocket infection, followed by pacemaker-induced cardiomyopathy, cardiac arrest that requires resuscitation, and upper limb DVT.

Studies that looked at the magnitude of complications related to pacemaker implantation found that the rate of complications ranged from 3.6% to 21.6%.8–17 Variations in patient sociodemographic factors among studies, patient volumes, and Centre-to-Centre differences in experience levels may all account for this heterogeneity in the rate of complications.

In comparison to studies conducted in Norway, Australia, Germany, the United States, Iraq, Senegal, Cameroon, and Ghana, which showed complication rates of 12%, 7.2%, 3%, 4.2%, 4.25%, 3.9%, 9.7%, and 7%, respectively, this study’s complication rate of 15.3% is greater.8–10,12,13,15,16 But compared to research from the Netherlands and Nigeria, where complication rates were reported to be 21.6% and 17.6%, respectively, our study’s complication rate was lower.11,14

Pneumothorax is a pacemaker-related event in 0.6–3.9% of patients, according to studies. Pneumothorax occurred in 2.54% of study participants in our study, which is about in line with other studies’ findings.8,15,16,18–22 When compared to individual studies, it is greater than those conducted in Taiwan (0.6%), the United States (1.3%), the United States (1.5%), Denmark (0.9%), the Netherlands (2.24%), Norway (2.8%), Cameroon (1.6%), and Senegal (0.6%), but lower than those made in Spain (3.9%).8,15,16,18–22 This variation in pneumothorax magnitude between studies may be explained by variations in professional experience among centres, variations of venous access for implantation, and the inclusion of patients with CRT and ICD in quantifying the pneumothorax rate in some of these researches.

Pacemaker pocket Infection rate was 2.54% in our study, which is greater than those in Spain (1.61%), Iraq (0.25%), India (1.5%), Kenya (1.4%), but lower than Cameroon (3.1%).9,16,20,23,24 The variability in pocket infection rates between centres may be explained by the existence of a chronic illness, the patient’s immune system, and the use of different aseptic procedures.

In this study, lead dislodgment occurred in 2.5% of patients; this rate was higher than that of the United States (2.4%), Poland (2.4%), the Netherlands (4.9%), Iraq (0.25%), and Senegal (0.6%), but lower than that of Kenya.9,11,13,15,23,25 The use of various fixation types and differences in follow-up duration in the studies can account for this disparity in lead dislodgement rate.

This study found 1.7% pacemaker-induced cardiomyopathies, which is less than studies from Korea (16.1%), Norway (9%), and the United States (12.3%).26–28 The lower rate of pacemaker-related cardiomyopathy in our study can be attributed to the shorter follow-up time.

We reported an upper limb DVT rate of 1.7%, which was lower than studies conducted in Brazil (6%) and Iran (21%), but was greater than the one from Australia (0.15%).10,29,30 Prior reports used different imaging modality approaches, and types of cardiac rhythm management devices, which could be the cause of this significant variation in studies results.

Our studies’ 1.7% pacemaker-related cardiac arrest necessitating resuscitation rate is significantly lower than studies conducted in 14 sub-Saharan countries, which found that 20% pacemaker-related cardiac arrest required resuscitation.31 In the sub-Saharan study done in 14 countries, there were more pacemaker-rated cardiac arrests during the skill transfer to cardiologists by the trainers.

In our study, 0.8% of patients developed a pocket hematoma, which is higher than studies conducted in Germany (0.2%) and Spain (0.32%), but lower than studies conducted in Cameroon (1.5%), Iraq (1%), and the Netherlands (2.9%).9,11,12,16,21 This discrepancy in pocket hematoma rate can be explained by differences in age distribution and comorbidities like coagulopathy.

Our study’s early battery failure rate of 0.8% is greater than the global rate of 0.24%.32 The advancement in pacemaker batteries can be attributed to the generally lower rate of early battery depletion.

The rate of mortality following pacemaker implantation primarily depends on the length of follow-up and may or may not be related to the pacemaker. Our study found a death rate that was higher than studies conducted in Australia (0.73%), Iraq (0%), the USA (0.08%), and Kenya (0.0%), but lower than those conducted in Senegal (1.2%), Spain (5.2%), and France (3.9%).9,10,13,15,20,23,33

Studies conducted in the USA and Australia that were similar to ours showed an association between sex and complications related to pacemakers.10,22 Because they have smaller veins, thin vessel walls, a smaller right ventricle, and less tissue between the subclavian vein and pleura, females are more likely to experience complications from pacemakers.

Similar to our research results, studies from the USA, and Netherlands, have shown that complications related to pacemakers are associated with aging. This is because older people have more comorbidities than younger people.11,34 In contrast, a study conducted in Turkey revealed that younger patients have a greater rate of complications associated with pacemakers.35

As with any procedure, operator experience is essential to reducing pacemaker-related complications, as was found in our study. This also received backing from research from the United States and Norway.8,13

In our facility, one of the follow-up procedures is evaluating pacemaker abnormalities to identify them early and manage them to avoid complications. However, just 45.8% of our patients were adherent to follow-up, which is the biggest problem with our patients. An issue with adherence was found by various other centres as well. For example, two studies from the USA revealed that only 42.4% and 41.3% of patients adhered to the initial 2–12 weeks of follow-up.36

Conclusion

Pacemaker implantation is associated with many complications. Pneumothorax, pacemaker pocket infection, and lead dislodgement were the most common complications. Patient age at pacemaker implantation, patient gender, implant group type, and follow-up compliance were factors associated with pacemaker-related complications. Skill development through specialized training and compliance counseling may improve outcomes for patients who have complications related to implant group type and poor adherence to follow.

Limitations of the Study

This study had limitations. First, because this was a single-center study with a small sample size, it is difficult to generalize. Second, the retrospective nature of the study may have introduced classification bias into this study. However, it provides valuable information for quality improvement by providing baseline complication data and data for comparing pacemaker complications from other centers and pointing out areas of concern and areas that need improvement to reduce complications.

Data Sharing Statement

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Consent and Ethical Clearance

The study complied with the Declaration of Helsinki. A letter of ethical clearance was obtained from Addis Ababa University, College of Health Sciences, Institutional Review Board (protocol number =013/22SNM). Written informed consent was obtained from each patient to participate in the interview and to extract data from their medical charts. Privacy and confidentiality were ensured during patient interview and medical chart review.

Acknowledgment

We would like to acknowledge the participants and data collectors for providing us consent to share his history and collecting data respectively. We also want to mention our gratitude to Addis Ababa University for giving us ethical clearance.

Funding

No specific funding for publication of this paper.

Disclosure

The authors declare that there is no conflict of interest regarding the publication of this article.

References

1. Coombes D. Pacemaker therapy 1: clinical indications, placement and complications. Nursing Times. 2021;117:11–22.

2. Glikson M, Nielsen JC, Kronborg MB, et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: developed by the task force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) With the special contribution of the European Heart Rhythm Association (EHRA). EP Europace. 2022;24(1):71–164.

3. Cantillon DJ, Exner DV, Badie N, et al. Complications and health care costs associated with transvenous cardiac pacemakers in a nationwide assessment. JACC. 2017;3(11):1296–1305. doi:10.1016/j.jacep.2017.05.007

4. Palmer SJ. Post-implantation pacemaker complications: the nurse’s role in management. Br J Cardiac Nurs. 2014;9(12):592–598. doi:10.12968/bjca.2014.9.12.592

5. Gugu KM. Early complications after pacemaker implantations. In: InCardiac Pacemakers-Biological Aspects, Clinical Applications and Possible Complications. IntechOpen; 2011.

6. Schiariti M, Cacciola M, Puddu PE. Complications of pacemaker implantation. Res Gate. 2011;38(2):271–298.

7. Stevenson RT, Lugg D, Gray R, Hollis D, Stoner M, Williams JL. Pacemaker implantation in the extreme elderly. J Interv Card Electrophysiol. 2012;33(1):51–58. doi:10.1007/s10840-011-9618-5

8. Haug B, Kjelsberg K, Lappegård KT. Pacemaker implantation in small hospitals: complication rates comparable to larger centres. Europace. 2011;13(11):1580–1586. doi:10.1093/europace/eur162

9. Ahmed MH. Patients characteristic, indications, and complications of permanent pacemaker implantation: a prospective single-center study. Med J Babylon. 2022;19(2):152. doi:10.4103/MJBL.MJBL_3_22

10. Vijayarajan V, Kritharides L, Brieger D, Cheng YY, Chow V, Ng AC. Sex differences in rates of permanent pacemaker implantation and in-hospital complications: a statewide cohort study of over 7 million persons from 2009–2018. PLoS One. 2022;17(8):e0272305. doi:10.1371/journal.pone.0272305

11. Udo EO, Zuithoff NP, van Hemel NM, et al. Incidence and predictors of short-and long-term complications in pacemaker therapy: the FOLLOWPACE study. Heart Rhythm. 2012;9(5):728–735. doi:10.1016/j.hrthm.2011.12.014

12. Hasan F, Nedios S, Karosiene Z, et al. Perioperative complications after pacemaker implantation: higher complication rates with subclavian vein puncture than with cephalic vein cutdown. J Interv Card Electrophysiol. 2023;66(4):857–863. doi:10.1007/s10840-022-01135-x

13. Tobin K, Stewart J, Westveer D, Frumin H. Acute complications of permanent pacemaker implantation: their financial implication and relation to volume and operator experience. Am J Cardiol. 2000;85(6):774–776. doi:10.1016/S0002-9149(99)00861-9

14. Falase B, Sanusi M, Johnson A. Analysis of a five year experience of permanent pacemaker implantation at a Nigerian Teaching Hospital: need for a national database. Pan Afr Med J. 2014;16:1.

15. Kane A, Sarr SA, Ndobo JV, et al. Cardiac pacing challenge in Sub-Saharan Africa environnement: experience of the cardiology department of teaching Hospital Aristide Le Dantec in Dakar. BMC Cardiovascul Disorder. 2019;19(1):1–7. doi:10.1186/s12872-019-1176-2

16. Tchoumi JC. Permanent cardiac pacing activities in a tertiary sub-Saharan centre. J Xiangya Med. 2017;2:8.

17. Tettey M, Entsua-Mensah K, Ernest A, Ofosu-Appiah E. Permanent pacemaker implantation in Accra, an 18-year review. Ann Afr Chir Thor Cardiovasc. 2012;7:2.

18. Lin YS, Hung SP, Chen PR, et al. Risk factors influencing complications of cardiac implantable electronic device implantation: infection, pneumothorax and heart perforation: a nationwide population-based cohort study. Medicine. 2014;93(27):e213. doi:10.1097/MD.0000000000000213

19. Ogunbayo GO, Charnigo R, Darrat Y, et al. Incidence, predictors, and outcomes associated with pneumothorax during cardiac electronic device implantation: a 16-year review in over 3.7 million patients. Heart Rhythm. 2017;14(12):1764–1770. doi:10.1016/j.hrthm.2017.07.024

20. Kirkfeldt RE, Johansen JB, Nohr EA, Moller M, Arnsbo P, Nielsen JC. Pneumothorax in cardiac pacing: a population-based cohort study of 28 860 Danish patients. Europace. 2012;14(8):1132–1138. doi:10.1093/europace/eus054

21. Carrión-Camacho MR, Marín-León I, Molina-Doñoro JM, González-López JR. Safety of permanent pacemaker implantation: a prospective study. J Clin Med. 2019;8(1):35. doi:10.3390/jcm8010035

22. Ellenbogen KA, Hellkamp AS, Wilkoff BL, et al. Complications arising after implantation of DDD pacemakers: the MOST experience. Am J Cardiol. 2003;92(6):740–741. doi:10.1016/S0002-9149(03)00844-0

23. Juma PV Baseline Profile, Rhythm Abnormalities and Outcomes for Patients With Pacemakers Inserted at the Kenyatta National Hospital [Doctoral dissertation]. University of Nairobi; 2017.

24. Datta G. Pacemaker pocket infection rate and suture technique. Turk Kardiyoloji Dernegi Arsivi. 2020;48(1):44. doi:10.5543/tkda.2019.82598

25. Dębski M, Ulman M, Ząbek A, et al. Lead-related complications after DDD pacemaker implantation. Kardiologia Polska. 2018;76(8):1224–1231. doi:10.5603/KP.a2018.0089

26. Kim JH, Kang KW, Chin JY, Kim TS, Park JH, Choi YJ. Major determinant of the occurrence of pacing-induced cardiomyopathy in complete atrioventricular block: a multicentre, retrospective analysis over a 15-year period in South Korea. BMJ open. 2018;8(2):e019048. doi:10.1136/bmjopen-2017-019048

27. Dreger H, Maethner K, Bondke H, Baumann G, Melzer C. Pacing-induced cardiomyopathy in patients with right ventricular stimulation for> 15 years. Europace. 2012;14(2):238–242. doi:10.1093/europace/eur258

28. Kiehl EL, Makki T, Kumar R, et al. Incidence and predictors of right ventricular pacing-induced cardiomyopathy in patients with complete atrioventricular block and preserved left ventricular systolic function. Heart Rhythm. 2016;13(12):2272–2278. doi:10.1016/j.hrthm.2016.09.027

29. Neto AS, Costa R, Caldas JG, Filho MM, Filho MM. Incidence and risk factors of upper extremity deep vein lesions after permanent transvenous pacemaker implant: a 6‐month follow‐up prospective study. Pacing Clin Electrophysiol. 2002;25(9):1301–1306. doi:10.1046/j.1460-9592.2002.01301.x

30. Safi M, Akbarzadeh MA, Azinfar A, Namazi MH, Khaheshi I. Upper extremity deep venous thrombosis and stenosis after implantation of pacemakers and defibrillators; A prospective study. Rom J Intern Med. 2017;55(3):139–144. doi:10.1515/rjim-2017-0018

31. Jouven X, Diop BI, Narayanan K, et al. Cardiac Pacing in Sub-Saharan Africa: JACC International. J Am Coll Cardiol. 2019;74(21):2652–2660. doi:10.1016/j.jacc.2019.09.034

32. Bhargava K, Arora V, Jaswal A, Vora A. Premature battery depletion with St. Jude medical ICD and CRT-D devices: Indian heart rhythm society guidelines for physicians. Indian Heart J. 2019;71(1):12–14. doi:10.1016/j.ihj.2016.12.014

33. Villalba S, Roda J, Quesada A, et al. Retrospective study of patients who undergo pacemaker implantation in short-stay ambulatory surgery. Long-term follow-up and cost analysis. Revista Española de Cardiología. 2004;57(3):234–240. doi:10.1016/S0300-8932(04)77095-5

34. Armaganijan LV, Toff WD, Nielsen JC, et al. Are elderly patients at increased risk of complications following pacemaker implantation? A meta‐analysis of randomized trials. Pacing Clin Electrophysiol. 2012;35(2):131–134. doi:10.1111/j.1540-8159.2011.03240.x

35. Özcan KS, Osmonov D, Altay S, et al. Pacemaker implantation complication rates in elderly and young patients. Clin Intervent Aging. 2013;7:1051–1054. doi:10.2147/CIA.S47121

36. Hess PL, Mi X, Curtis LH, Wilkoff BL, Hegland DD, Al-Khatib SM. Follow-up of patients with new cardiovascular implantable electronic devices: is adherence to the experts’ recommendations associated with improved outcomes? Heart Rhythm. 2013;10(8):1. doi:10.1016/j.hrthm.2013.06.009

Creative Commons License © 2024 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.