Wear studies on plasma-sprayed Al2O3 and 8mole% of Yttrium-stabilized ZrO2 composite coating on biomedical Ti-6Al-4V alloy for orthopedic joint application

Authors Ganapathy P, Manivasagam G, Rajamanickam A, Natarajan A

Received 19 January 2015

Accepted for publication 18 March 2015

Published 1 October 2015 Volume 2015:10(Supplement 1 Challenges in biomaterials research) Pages 213—222

DOI https://doi.org/10.2147/IJN.S79997

Checked for plagiarism Yes

Review by Single-blind

Peer reviewer comments 3

Editor who approved publication: Dr Thomas J Webster


Perumal Ganapathy,1 Geetha Manivasagam,2 Asokamani Rajamanickam,3 Alagumurthi Natarajan4

1Department of Mechanical Engineering, VRS College of Engineering and Technology, Arasur, Villupuram, Tamil Nadu, India; 2Centre for Biomaterials Science and Technology, SMBS, VIT University, Vellore, Tamil Nadu, India; 3Department of Physics, Dhanalakshmi Engineering College, Anna University, Chennai, Tamil Nadu, India; 4Department of Mechanical Engineering, Pondicherry Engineering College, Pondicherry, Puducherry, India

Abstract: This paper presents the wear characteristics of the composite ceramic coating made with Al2O3-40wt%8YSZ on the biomedical grade Ti-6Al-4V alloy (grade 5) used for total joint prosthetic components, with the aim of improving their tribological behavior. The coatings were deposited using a plasma spraying technique, and optimization of plasma parameters was performed using response surface methodology to obtain dense coating. The tribological behaviors of the coated and uncoated substrates were evaluated using a ball-on-plate sliding wear tester at 37°C in simulated body-fluid conditions. The microstructure of both the titanium alloy and coated specimen were examined using an optical microscope and scanning electron microscope. The hardness of the plasma-sprayed alumina–zirconia composite coatings was 2.5 times higher than that of the Ti-6Al-4V alloy, while the wear rate of Ti-6Al-4V alloy was 253 times higher than that of the composite-coated Ti-6Al-4V alloy. The superior wear resistance of the alumina–zirconia coated alloy is attributed to its enhanced hardness and intersplat bonding strength. Wear-track examination showed that the predominant wear mechanism of Ti-6Al-4V alloy was abrasive and adhesive wear, whereas, in the case of alumina–zirconia composite coated alloy, the wear was dominated by microchipping and microcracking.

Keywords: Ti-6Al-4V alloy, alumina–zirconia, wear, Hank’s solution, titanium

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.