Comparisons of amplitude of pseudoaccommodation with aspheric yellow, spheric yellow, and spheric clear monofocal intraocular lenses

Authors Nishi T, Taketani F, Ueda T, Ogata N

Received 10 August 2013

Accepted for publication 17 September 2013

Published 1 November 2013 Volume 2013:7 Pages 2159—2164

DOI https://doi.org/10.2147/OPTH.S52771

Checked for plagiarism Yes

Review by Single-blind

Peer reviewer comments 2

Tomo Nishi, Futoshi Taketani, Tetsuo Ueda, Nahoko Ogata

Department of Ophthalmology, Nara Medical University, Kashihara City, Nara, Japan

Purpose: To determine the amplitude of pseudoaccommodation and higher-order aberrations with three types of implanted monofocal intraocular lenses (IOLs): aspheric yellow (IQ); spheric yellow (NT); and spheric clear (AT).
Setting: Department of Ophthalmology, Nara Medical University, Nara, Japan.
Methods: We studied 60 patients who underwent small incision phacoemulsification with the implantation of a monofocal IQ, NT, or AT IOL. The pseudoaccommodation was measured by the lens-loading method, and the postoperative ocular higher-order aberrations were measured with a Hartmann–Shack wavefront analyzer through natural and 4 mm pupils.
Results: Sixty eyes of 60 patients were studied. The average amplitude of the pseudoaccommodation was 0.45±0.24 D with the IQ IOL, which was significantly lower than that with the AT IOL at 0.81±0.37 D (Tukey's test; P<0.01). The differences in the amplitude of the pseudoaccommodation between the IQ and the NT IOLs, and between the NT and the AT IOLs were not significant (Tukey's test; P>0.05). The degree of spherical aberration was significantly different for the IQ, NT, and AT lenses (analysis of variance, P=0.016). The spherical aberration through the IQ IOL was significantly lower than that through the NT and the AT IOLs (Tukey's test; P<0.01). The fourth-order RMS (root mean square) aberration of the IQ lens was also significantly lower than that of the NT and AT IOLs (Tukey's test; P<0.01).
Conclusion: Our results suggest that the spherical aberration and selective spectral transmission of IOLs may work together to increase the amplitude of the pseudoaccommodation.

Keywords: pseudoaccommodation, aspheric intraocular lens, spheric intraocular lens, higher-order aberrations

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.

Download Article [PDF]  View Full Text [HTML][Machine readable]

 

Other article by this author:

Readers of this article also read:

Green synthesis of water-soluble nontoxic polymeric nanocomposites containing silver nanoparticles

Prozorova GF, Pozdnyakov AS, Kuznetsova NP, Korzhova SA, Emel’yanov AI, Ermakova TG, Fadeeva TV, Sosedova LM

International Journal of Nanomedicine 2014, 9:1883-1889

Published Date: 16 April 2014

Methacrylic-based nanogels for the pH-sensitive delivery of 5-Fluorouracil in the colon

Ashwanikumar N, Kumar NA, Nair SA, Kumar GS

International Journal of Nanomedicine 2012, 7:5769-5779

Published Date: 15 November 2012

A novel preparation method for silicone oil nanoemulsions and its application for coating hair with silicone

Hu Z, Liao M, Chen Y, Cai Y, Meng L, Liu Y, Lv N, Liu Z, Yuan W

International Journal of Nanomedicine 2012, 7:5719-5724

Published Date: 12 November 2012

Cross-linked acrylic hydrogel for the controlled delivery of hydrophobic drugs in cancer therapy

Deepa G, Thulasidasan AK, Anto RJ, Pillai JJ, Kumar GS

International Journal of Nanomedicine 2012, 7:4077-4088

Published Date: 27 July 2012

Corrigendum: Softec HD hydrophilic acrylic intraocular lens: biocompatibility and precision

Espandar L, Sikder S, Moshirfar M

Clinical Ophthalmology 2011, 5:159-160

Published Date: 6 February 2011

Crystallization after intravitreal ganciclovir injection

Pitipol Choopong, Nattaporn Tesavibul, Nattawut Rodanant

Clinical Ophthalmology 2010, 4:709-711

Published Date: 14 July 2010