Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer

Authors Farhood QK

Received 22 September 2012

Accepted for publication 6 November 2012

Published 27 December 2012 Volume 2013:7 Pages 23—27

DOI https://doi.org/10.2147/OPTH.S38418

Checked for plagiarism Yes

Review by Single-blind

Peer reviewer comments 2

Qasim K Farhood

Department of Ophthalmology, College of Medicine, University of Babylon, Babylon, Iraq

Purpose: Tonometry, or measurement of intraocular pressure (IOP), is one of the most important examination procedures in ophthalmic clinics, and IOP is an important parameter in the diagnosis of glaucoma. Because there are numerous types of tonometer available, it is important to evaluate the differences in readings between different tonometers. Goldmann applanation tonometers (GATs) and noncontact air-puff tonometers (APTs) are largely available in ophthalmic clinics. The purpose of this study was to evaluate the role of AP tonometer by comparing the measurements of IOP made using this device with those made using a GAT.
Patients and methods: This study involved 196 eyes from 98 study participants, all of whom were patients attending an ophthalmic outpatient clinic. Each patient's IOP was measured using both Goldmann applanation tonometry and AP tonometry, and the difference in readings between the two methods was calculated.
Results: The mean IOP as measured by GAT was 13.06 ± 4.774 mmHg, while that as measured by AP tonometer was 15.91 ± 6.955 mmHg. The mean difference between the two methods of measurement was 2.72 ± 2.34 mmHg. The readings obtained by AP tonometer were higher than those obtained by GAT in 74% of patients, and this difference was most obvious when the GAT measurement of IOP exceeded 24 mmHg. No statistically significant variation in IOP was noted between the devices when the patients' age, sex, and laterality (right and left eyes) were considered.
Conclusion: There is a significant difference in the measurement of IOP between GATs and AP tonometers. Goldmann applanation tonometry remains the most suitable and reliable method for measuring IOP. Because measurements of IOP by AP tonometer are usually higher than those obtained by GAT regardless of the patient's age, sex, or laterality of eyes, AP tonometry is a suitable method for community or mass screenings of IOP.

Keywords: tonometry, comparison, glaucoma, noncontact tonometry, goldmann applanation tonometer

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.

Download Article [PDF]  View Full Text [HTML][Machine readable]

 

Other articles by this author:

Readers of this article also read:

Magnetic microparticle-based multimer detection system for the detection of prion oligomers in sheep

Lim K, Kim SY, Lee B, Segarra C, Kang S, Ju YR, Schmerr MJ, Coste J, Kim SY, Yokoyama T, An SSA

International Journal of Nanomedicine 2015, 10:241-250

Published Date: 9 September 2015

Emerging and future therapies for hemophilia

Carr ME, Tortella BJ

Journal of Blood Medicine 2015, 6:245-255

Published Date: 3 September 2015

Age-associated decrease in global DNA methylation in patients with major depression

Tseng PT, Lin PY, Lee Y, Hung CF, Lung FW, Chen CS, Chong MY

Neuropsychiatric Disease and Treatment 2014, 10:2105-2114

Published Date: 10 November 2014

Can low brain-derived neurotrophic factor levels be a marker of the presence of depression in obese women?

Celik Guzel E, Bakkal E, Guzel S, Eroglu HE, Acar A, Kuçukyalcin V, Topcu B

Neuropsychiatric Disease and Treatment 2014, 10:2079-2086

Published Date: 5 November 2014

A novel preparation method for silicone oil nanoemulsions and its application for coating hair with silicone

Hu Z, Liao M, Chen Y, Cai Y, Meng L, Liu Y, Lv N, Liu Z, Yuan W

International Journal of Nanomedicine 2012, 7:5719-5724

Published Date: 12 November 2012

Crystallization after intravitreal ganciclovir injection

Pitipol Choopong, Nattaporn Tesavibul, Nattawut Rodanant

Clinical Ophthalmology 2010, 4:709-711

Published Date: 14 July 2010