Back to Journals » International Journal of Chronic Obstructive Pulmonary Disease » Volume 7

Total lung capacity by plethysmography and high-resolution computed tomography in COPD

Authors Garfield J, Marchetti N, Gaughan JP, Steiner, Criner G

Received 22 September 2011

Accepted for publication 19 November 2011

Published 22 February 2012 Volume 2012:7 Pages 119—126

DOI https://doi.org/10.2147/COPD.S26419

Review by Single anonymous peer review

Peer reviewer comments 2



Jamie L Garfield, Nathaniel Marchetti, John P Gaughan, Robert M Steiner, Gerard J Criner

Department of Pulmonary and Critical Care Medicine and Department of Radiology, Temple University School of Medicine, Philadelphia, Pennsylvania, USA

Aim: To characterize and compare total lung capacity (TLC) measured by plethysmography with high-resolution computed tomography (HRCT), and to identify variables that predict the difference between the two modalities.
Methods: Fifty-nine consecutive patients referred for the evaluation of COPD were retrospectively reviewed. Patients underwent full pulmonary function testing and HRCT within 3 months. TLC was obtained by plethysmography as per American Thoracic Society/European Respiratory Society standards and by HRCT using custom software on 0.75 and 5 mm thick contiguous slices performed at full inspiration (TLC).
Results: TLC measured by plethysmography correlated with TLC measured by inspiratory HRCT (r = 0.92, P < 0.01). TLC measured by plethysmography was larger than that determined by inspiratory HRCT in most patients (mean of 6.46 ± 1.28 L and 5.34 ± 1.20 L respectively, P < 0.05). TLC measured by both plethysmography and HRCT correlated significantly with indices of airflow obstruction (forced expiratory volume in 1 second/forced vital capacity [FVC] and FVC%), static lung volumes (residual volume, percent predicted [RV%], total lung capacity, percent predicted [TLC%], functional residual capacity, percent predicted [FRC%], and inspiratory capacity, percent predicted), and percent emphysema. TLC by plethysmography and HRCT both demonstrated significant inverse correlations with diffusion impairment. The absolute difference between TLC measured by plethysmography and HRCT increased as RV%, TLC%, and FRC% increased. Gas trapping (RV% and FRC%) independently predicted the difference in TLC between plethysmography and HRCT.
Conclusion: In COPD, TLC by plethysmography can be up to 2 L greater than inspiratory HRCT. Gas trapping independently predicts patients for whom TLC by plethysmography differs significantly from HRCT.

Keywords: lung capacity, plethysmography, high-resolution computed tomography, gas trapping, lung volume measurement errors

Creative Commons License © 2012 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.