Technical and clinical validation of an environmental exposure unit for ragweed

Authors Gomes PJ, Lane KJ, Angjeli E, Stein L, Abelson MB

Received 30 September 2016

Accepted for publication 11 November 2016

Published 14 December 2016 Volume 2016:9 Pages 215—221

DOI https://doi.org/10.2147/JAA.S123547

Checked for plagiarism Yes

Review by Single-blind

Peer reviewers approved by Dr Lucy Goodman

Peer reviewer comments 3

Editor who approved publication: Dr Luis Garcia-Marcos


Paul J Gomes,1 Keith J Lane,1 Endri Angjeli,1 Linda Stein,1 Mark B Abelson1,2

1Ora, Inc., Andover, MA, USA; 2Department of Ophthalmology, Harvard Medical School, Cambridge, MA, USA

Background: Allergic rhinitis is a common condition, with ragweed pollen one of the more prevalent aeroallergens. Environmental exposure units such as the Allergen BioCube® are valuable models for clinical allergy studies. A study was conducted to validate the Allergen BioCube for uniform ragweed pollen concentrations and clinically relevant sign and symptom responses to ragweed exposure.
Methods: Ragweed pollen concentrations were measured on 3 consecutive days in the Allergen BioCube and verified by Rotorod collection and continuous laser particle count measurements. Subjects (N=10) were exposed to ragweed pollen in the BioCube for 3 hours per day for 3 consecutive days. Subjects assessed their nasal itching, sneezing, rhinorrhea, and nasal congestion during each BioCube exposure; total nasal symptom score was computed. Peak nasal inspiratory flow was also assessed during BioCube exposure.
Results: Uniform ragweed pollen concentrations were obtained throughout each of the 3-hour testing periods in the Allergen BioCube, both spatially and temporally, at all subject positions, with a low mean standard deviation of 10%. Pronounced increases in mean total nasal symptom scores (6.7±0.94 to 7.6±0.86, last 90 minutes of exposure) occurred for all three BioCube ragweed pollen exposure visits. Mean peak nasal inspiratory flow decreased 24% at 3 hours of BioCube exposure on Day 3. No safety issues of concern occurred in this study.
Conclusion: The Allergen BioCube achieved technical and clinical validation for ragweed allergen. Ragweed pollen concentration was uniform both temporally and spatially. Allergic rhinitis signs and symptoms were induced in subjects during exposure to ragweed in the BioCube at clinically meaningful levels for allergy studies.

Keywords: Allergen BioCube®, validation, ragweed, symptom response, exposure unit

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.