Back to Journals » Medical Devices: Evidence and Research » Volume 8

Randomized trial comparing episiotomies with Braun-Stadler episiotomy scissors and EPISCISSORS-60®

Authors Sawant G, Kumar D

Received 23 February 2015

Accepted for publication 2 April 2015

Published 1 June 2015 Volume 2015:8 Pages 251—254

DOI https://doi.org/10.2147/MDER.S83360

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 1

Editor who approved publication: Dr Scott Fraser


Ganpat Sawant, Divya Kumar

Dr DY Patil Medical College and Hospital, Nerul, Navi Mumbai, India

Introduction: Episiotomy angle is a crucial factor in causation of obstetric anal sphincter injuries (OASIS), which are the major cause of female bowel incontinence. Sutured episiotomies angled too close to the midline (<30 degree) or too far away from the midline (.60 degree) fail to unload the perineum sufficiently and predispose to OASIS. A 25-degree post-delivery episiotomy suture angle has a 10% risk of OASIS while 45-degree episiotomy is associated with 0.5% risk. To account for perineal distension at crowning, a 60-degree episiotomy incision is required to achieve 43–50 degree suture angles. We compared episiotomy suture angles with commonly used Braun-Stadler episiotomy scissors with the new fixed angle EPISCISSORS-60®.
Methods: Ethical approval was obtained. A prospective cluster randomization design was chosen. Thirty-one patients were required in each group for a 12-degree difference with power at 90% and 5% significance. Sutured episiotomy angles and post-delivery linear distance from caudal end of the sutured episiotomy to the anus were measured with protractors and rulers. Two-tailed t-tests were used to compare the two groups.
Results: Thirty-one nulliparae had episiotomies with EPISCISSORS-60®, 32 with Braun-Stadler. Mean age (25 versus 24.8 years) was similar. EPISCISSORS-60® episiotomies were angled 12 degrees more laterally away from the anus compared to Braun-Stadler (40.6 degrees, 95% confidence interval [CI] ±2, interquartile range [IQR] 35–45 versus 28.3 degrees, 95% CI ±2, IQR 25–30, P<0.0001). The post-delivery linear distance from caudal end of the sutured episiotomy to the anus was 15 mm more with the EPISCISSORS-60® compared to Braun-Stadler (35 mm, 95% CI ±2.2, IQR =30–39 versus 19.5; 95% CI ±1.3, IQR =14.75–22.25 P<0.0001). EPISCISSORS-60® episiotomies measured longer (47 mm versus 40 mm, P<0.0001). There were no OASIS cases in the EPISCISSORS-60® group versus one in the Braun-Stadler group.
Conclusion: The EPISCISSORS-60® sutured episiotomies are much further away from the midline in angular and distance measures, hence at lower OASIS risk.

Keywords: episiotomy, EPISCISSORS-60®, obstetric anal sphincter injuries (OASIS), perineal tears
 

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.