Five-year extension of a clinical trial comparing the EX-PRESS glaucoma filtration device and trabeculectomy in primary open-angle glaucoma

Authors de Jong, Lafuma A, Aguade, Berdeaux G

Published 29 April 2011 Volume 2011:5 Pages 527—533

DOI https://doi.org/10.2147/OPTH.S18565

Review by Single anonymous peer review

Peer reviewer comments 2



Leo de Jong1, Antoine Lafuma2, Anne-Sophie Aguadé2, Gilles Berdeaux3,4
1Academic Medical Centre, Ophthalmology Department, Amsterdam, The Netherlands, 2Cemka-Eval, Bourg la Reine, Paris, France, 3Alcon France, Rueil-Malmaison, Paris, France, 4Conservatoire National des Arts et Métiers, Paris, France

Background: This study compared the efficacy of the EX-PRESS® glaucoma filtration device and trabeculectomy in primary open-angle glaucoma up to five years after surgery.
Methods: Patients from a previously reported randomized, open-label, parallel-arm clinical trial in which 78 patients received either the EX-PRESS glaucoma filtration device or underwent a trabeculectomy were followed for up to an additional four years (five total) beyond the original study (39 eyes per treatment group). Risk-benefit data were obtained for up to five years after glaucoma surgery. Outcome variables were intraocular pressures and intraocular pressure medications. Complete success was denoted by intraocular pressure values ≤ 18 mmHg without medication.
Results: The EX-PRESS glaucoma filtration device controlled intraocular pressure more effectively without medication for more patients from year 1 (86.8% versus 61.5%, P = 0.01) to year 3 (66.7% versus 41.0%, P = 0.02) than trabeculectomy. At year 1, only 12.8% of patients required intraocular pressure medication after EX-PRESS implantation, compared with 35.9% after trabeculectomy. The proportions became closer at year 5 (41% versus 53.9%). The responder rate was higher with EX-PRESS and time to failure was longer. In addition, surgical interventions for complications were fewer after EX-PRESS implantation.
Conclusion: This five-year analysis confirmed and extended the results reported after one year. Compared with trabeculectomy, EX-PRESS provided better intraocular pressure control in the first three years, and patients required fewer intraocular pressure medications and fewer surgical interventions during the five-year study period. For patients with primary open-angle glaucoma, the EX-PRESS glaucoma filtration device, implanted under a superficial scleral flap, produced significantly higher success rates than trabeculectomy. EX-PRESS is an effective device for long-term treatment of primary open-angle glaucoma.

Keywords: glaucoma filtration device, trabeculectomy, primary open-angle glaucoma, intraocular pressure, medication

Creative Commons License © 2011 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.