Biomechanical stability of a bioabsorbable self-retaining polylactic acid/nano-sized β-tricalcium phosphate cervical spine interbody fusion device in single-level anterior cervical discectomy and fusion sheep models

Authors Cao L, Duan, Li, Yuan, Zhao M, Che W, Wang, Dong J

Received 19 September 2012

Accepted for publication 30 October 2012

Published 27 November 2012 Volume 2012:7 Pages 5875—5880

DOI https://doi.org/10.2147/IJN.S38288

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 2



Lu Cao,1 Ping-Guo Duan,1 Xi-Lei Li,1 Feng-Lai Yuan,3 Ming-Dong Zhao,2 Wu Che,1 Hui-Ren Wang,1 Jian Dong1

1Department of Orthopedic Surgery, Zhongshan Hospital, State Key Laboratory of Molecular Engineering of Polymers, Fudan University, Shanghai, China; 2Department of Orthopedic Surgery, Jinshan Hospital, Fudan University, Shanghai, China; 3Affiliated Third Hospital of Nantong University, Wuxi, China

Purpose: The aim of this study was to investigate the biomechanical stability provided by a novel, polylactic acid/nano-sized, β-tricalcium phosphate, bioabsorbable, self-retaining cervical fusion cage (BCFC).
Methods: Quasistatic nonconstraining torques (maximum 1.5 NM) induced flexion, extension, lateral bending (±1.5 NM), and axial rotation (±1.5 NM) on 32 sheep cervical spines (C2–C5). The motion segment C3–C4 was first tested intact; the following groups were tested after complete discectomy: autologous tricortical iliac crest bone graft, Medtronic–Wego polyetheretherketone (PEEK) cage, Solis PEEK cage, and BCFC. The autologous bone graft group was tested with an anterior plate. The mean range of motion (ROM) was calculated from the load-displacement curves.
Results: BCFC significantly decreased ROM in lateral bending and axial rotation compared to other implants, and no significant difference in ROM between two types of PEEK cages and BCFC could be observed in flexion and extension. Anterior cervical plate (ACP) significantly decreased ROM in flexion and extension, but no significant difference in ROM between BCFC and bone graft plus ACP could be determined in lateral bending and axial rotation.
Conclusion: The BCFC device showed better stability to autologous tricortical iliac crest bone graft and PEEK cages in single-level anterior cervical discectomy and fusion models and thus may be a potential alternative to the current PEEK cages.

Keywords: biomechanics, cervical spine, cages, bioabsorbable, sheep

Creative Commons License © 2012 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.