Application of Second-Generation Swept-Source Anterior Segment-OCT in the Measurement of Marginal Reflex Distance-1 (MRD-1)

Authors Ogasawara K

Received 20 January 2020

Accepted for publication 19 February 2020

Published 3 March 2020 Volume 2020:14 Pages 635—642

DOI https://doi.org/10.2147/OPTH.S246731

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 3

Editor who approved publication: Dr Scott Fraser



Kosuke Ogasawara

Medical Corporation, Ogasawara Eye Clinic, Morioka Iwate 020-0114, Japan

Correspondence: Kosuke Ogasawara
Ogasawara Eye Clinic, 3-10-12, Takamatsu, Morioka Iwate 020-0114, Japan
Tel +81 19 662 3223
Fax +81 19 662 7331
Email [email protected]

Purpose: To measure marginal reflex distance-1 (MRD-1) values in normal subjects and patients with blepharoptosis using second-generation swept-source anterior segment optical coherence tomography (AS-OCT) and to evaluate the usefulness of this measurement method.
Materials and Methods: MRD-1 values were measured using AS-OCT in 80 eyes of 40 normal subjects (16 men and 24 women) aged 20– 59 years with no neurological or endocrinological disorders, and the measured values were compared between four age groups comprising 20 eyes of 10 subjects each in their 20s, 30s, 40s, and 50s. The Volk Eye Check system was simultaneously used to measure MRD-1 in 40 eyes of 20 normal subjects, and the measured values were compared with those obtained using AS-OCT. In 21 eyes of consecutive 13 involutional blepharoptosis patients (48– 84 years of age, mean 70.2 years), MRD-1 values were measured using AS-OCT preoperatively and 3 months after surgery to examine the rates of successful measurements as well as preoperative and postoperative results.
Results: The mean MRD-1 value using AS-OCT in 80 eyes of 40 normal subjects was 3.70 ± 0.10 mm, whereas the values in 20 eyes of 10 subjects each from the four different age groups were 4.01 ± 0.54 mm, 3.88 ± 0.60 mm, 3.96 ± 0.77 mm, and 2.94 ± 0.79 mm for subjects in their 20s, 30s, 40s, and 50s, respectively; significant differences were observed between the 50s group and other groups (p< 0.01). The mean MRD-1 values using AS-OCT and the Volk Eye Check system in 40 eyes of 20 normal subjects were 3.68 ± 0.70 mm and 3.71 ± 0.84 mm, respectively, with no significant differences (p:0.87). The mean preoperative and postoperative MRD-1 values in patients with involutional blepharoptosis were 0.29 ± 0.70 mm and 3.24 ± 0.92 mm, respectively; the rate of successful MRD-1 measurements using AS-OCT was 100% among trial subjects.
Conclusion: MRD-1 measurement using AS-OCT is a simple and reproducible method that allows the direct comparison of numerical values in normal subjects as well as in patients with blepharoptosis.

Keywords: blepharoptosis, MRD1, swept-source anterior segment-OCT

Introduction

With increasing age of the population, more patients are developing blepharoptosis, thus increasing the need for its surgical treatment each year. Blepharoptosis causes aesthetic issues and visual function problems, especially narrowed visual field.15

International guidelines on the surgical treatment of blepharoptosis including preoperative evaluation and selection of the surgical procedure have not been established. Nevertheless, it is important to accurately measure the severity of blepharoptosis before surgery. According to the recent AAO opinion,4 blepharoptosis surgery is functionally indicated when marginal reflex distance 1 (MRD-1) is ≤2 mm. Common methods to measure MRD-1 include the corneal reflex-based measurement using a ruler (clinical method)6 and calculation using digital photographs and image J software program (http://rsbweb.nih.gov/ij; digital method).7 Recently, the Volk Eye Check system8 and analogous methods9,10 have also been reported. In this study, we used the second-generation swept-source anterior segment optical coherence tomography (AS-OCT) (CASIA2, Tomey Corporation, Nagoya, Japan) to measure MRD-1 values in normal subjects and preoperative and postoperative MRD-1 values in patients with blepharoptosis to assess its usefulness.

Materials and Methods

This study included adults with blepharoptosis or normal eyelid position. Written informed consent was obtained from all the subjects. The study was conducted as per the tenets of the Declaration of Helsinki and with the approval of the medical corporation of Ogasawara Eye Clinic’s ethical review board. Patients whose images and/or measurements appear in this publication provided additional written consent for publication of their photographs.

MRD-1 values were measured using AS-OCT in 80 eyes of 40 normal subjects (16 men and 24 women) aged 20–59 years who had no systemic diseases such as neurological and endocrinological disorders (Figure 1A). A snapshot taken during the measurement is shown in Figure 1B. MRD-1 values were measured as follows: (1) Images were separately taken for the left and right eyes in cataract preoperative measurement mode. Each subject placed his/her face on the chin support and fixed his/her eyes on a front orange fixation lamp with both eyes open naturally. (2) For analysis of MRD-1, an orthoptist clicked the front monitor and fixed the corneal reflex before adjusting the yellow line to 90°. (3) MRD-1 was automatically measured when the memory was placed on the superior palpebral margin and corneal apex. The measurements can be performed in a bright room. A typical example is shown in Figure 2. For patients whose superior palpebral margin was lower than the pupil center, the eyelid was elevated to confirm the pupil center and then returned to its original position before MRD-1 was measured (Figure 3).

Figure 1 MRD-1 measurement scene using AS-OCT (A). A snapshot taken during the MRD-1 measurement screen using AS-OCT (B).

Figure 2 A photograph of eyelids of a normal subject (A) and AS-OCT findings (B) A 26-year-old woman; MRD-1, right: 3.59 mm, left: 3.45 mm.

Figure 3 Photographs of eyelids of a patient with severe blepharoptosis (A), those being elevated to identify the pupil center (B), and those returned to the original position and subjected to MRD-1 measurement (C). A 84-year-old man; MRD-1, right: −0.23 mm, left: −0.58 mm.

The MRD-1 values measured using AS-OCT in normal subjects were compared to detect any significant difference between the four age groups comprising 20 eyes of 10 subjects each in their 20s, 30s, 40s, and 50s.

For comparison, the Volk Eye Check system8 was simultaneously used to measure MRD-1 in 40 eyes of 20 normal subjects aged 20–59 years of age, mean 39.2 years.

As a prospective study, preoperative and postoperative MRD-1 values were measured using AS-OCT in 21 eyes of consecutive 13 patients with involutional blepharoptosis (48–84 years of age, mean 68.6 years) to examine the rates of successful measurements, reproducibility, and values measured before and 3 months after surgery. The basic surgical procedure was a modified transconjunctival tucking method that was reported by us.11

In this study, t-test and analysis of variance (Bonferroni’s method) were used to detect significant differences.

Results

In 80 eyes of 40 normal subjects, the mean of MRD-1 values measured using AS-OCT was 3.70 ± 0.10 mm. In 20 eyes of 10 subjects each from the four different age groups, the mean MRD-1 values were 4.01 ± 0.54 mm, 3.88 ± 0.60 mm, 3.96 ± 0.77 mm, and 2.94 ± 0.79 mm for subjects in their 20s, 30s, 40s, and 50s, respectively; significant differences were observed between the 50s group and other groups (Figure 4). The rate of successful measurements was 100%, and the measurements were reproducible. Successful means MRD-1 was able to be measured definitely in trial subjects and get acceptable data.

Figure 4 MRD-1 measurement data from 20 eyes of 10 normal subjects in their 20s–50s. Significant differences were observed between the 50s group and other groups by analysis of variance (Bonferroni’s method). F (3,76)=10.60, p<0.01.

The mean of MRD-1 values measured using AS-OCT and Volk Eye Check system in 40 normal eyes of 20 subjects were 3.68 ± 0.70 mm and 3.71 ± 0.84 mm, respectively, with no significant differences between the two methods (Figure 5). The rate of successful measurements using AS-OCT was 100%, whereas that using the Volk Eye Check system,8 the result was 90%.

Figure 5 Results of MRD-1 measurement using AS-OCT and Volk Eye Check system in normal subjects. AS-OCT mean: 3.68 ± 0.70 mm, Volk Eye Check system mean: 3.71 ± 0.84 mm. No significant differences between the two methods were observed.

In patients with involutional blepharoptosis, MRD-1 values could be measured using AS-OCT for all cases, including severe ones. The mean MRD-1 values measured using AS-OCT before and 3 months after blepharoptosis surgery were 0.29 ± 0.70 mm and 3.24 ± 0.92 mm, respectively, demonstrating significant improvements (Figure 6). Photographs taken before and 3 months after surgery and for MRD-1 measurement for the representative patient are in shown Figures 3 and 7, respectively.

Figure 6 Preoperative (pre-ope) and postoperative (post-ope) MRD-1 measurements in patients included in this study. The mean MRD-1 values of pre-ope and 3 months post-ope were 0.29 ± 0.70 mm and 3.24 ± 0.92 mm, respectively.

Figure 7 A photograph taken 3 months after surgery for severe blepharoptosis (A) and MRD-1 measured using AS-OCT (B) A 84-year-old man:MRD-1, right: 2.70 mm, left: 3.21 mm. Before surgery photographs are shown in Figure 3.

Discussion

CASIA2 is an excellent instrument for AS-OCT, offering a variety of functions.12,13 These include cataract preoperative examination, gonioscopic analysis, and observation of the filtration bleb after glaucoma surgery, analysis of sagittal images of upper eyelid14 as well as corneal shape. In this study, we used AS-OCT functions to measure MRD-1 values in normal subjects and preoperative and postoperative MRD-1 values in patients with blepharoptosis to examine the usefulness of AS-OCT.

This study revealed that MRD-1 could be measured using AS-OCT in a simple manner, measured MRD-1 values could be directly compared as numerical values, and AS-OCT could be used to measure MRD-1 values in normal individuals as well as preoperative MRD-1 values in patients with severe blepharoptosis. In addition, because the reproducibility of measured values was high, it appears that AS-OCT is a clinically significant MRD-1 measurement method.

Compared with the AS-OCT method used in this study, previously reported methods for MRD-1 measurement have some drawbacks. The disadvantages of the clinical method include measurement errors as large as 0.5 mm6 and a complicated process required for MRD-1 analysis, although ImageJ is probably the most popular image analysis software program. The ORBSCAN2 topography system is used to improve the accuracy of values determined with the ImageJ program;15 however, this method is associated with several biases and has a disadvantage of indirectly measuring MRD-1 values. Choi et al9 compared MRD-1 values obtained from ImageJ analysis of photographs taken with four different cameras and reported MRD-1 errors varying approximately 0.5–0.7 mm depending on the differences in cameras used and positional and size differences of flashbulbs used.

In this study, we compared MRD-1 values that were measured using AS-OCT and those measured using the Volk Eye Check system.8 The Volk Eye Check System is a modified smartphone intended to collapse the space between convenience of clinical and the precision of digital measurements. This device is outfitted with image processing software that automatically measures parameters such as MRD-1 in real-time using photographs taken by integrated camera. No significant differences between the two methods were noted in normal subjects (Figure 5). However, in patients with blepharoptosis, MRD-1 could be measured using the Volk Eye Check system only for ≤50% of moderate to severe cases, and we could not compare the AS-OCT and Volk Eye Check system results in this population. Sinha et al16 compared the values measured using the clinical method, digital method, and Volk Eye Check system. Similarly to our results, they found no significant differences among the three groups in normal eyes. In patients with blepharoptosis, they found that the Volk Eye Check system overestimated the amount of blepharoptosis and the rate of successful measurements was approximately 66%. These and our data indicate advantage and superiority of AS-OCT compared to the Volk Eye Check System. Many Japanese individuals have thick eyelids that are shallow in depth and double eyelids are relatively minority. These characteristics may explain why software analysis of photographs is difficult (see a representative case in Figure 8). In AS-OCT measurement, there were concerns that the instrument might interfere with MRD-1 because it is placed right in front of the eyes. Our results indicate that this caused negligible errors, if any.

Figure 8 A photographs of eyelids of a 75-year-old male patient with severe blepharoptosis (A) and postoperative three months after surgery (B, C). Measuring MRD-1 values by AS-OCT (C) are 4.14 mm in right eye and 3.86 mm in left eye. Preoperative MRD-1, right eye:-0.81 mm, left eye:-1.19 mm.

The AS-OCT method allows simple MRD-1 measurement even in moderate and severe blepharoptosis patients with a stable measurable rate of 100%. In addition, this method may be useful to accurately and objectively assess the degree of recovery and postoperative course following blepharoptosis surgery. Furthermore, this method may be useful to identify unilateral blepharoptosis and compare the amounts of blepharoptosis between the left and right eyes.

Conclusions

MRD-1 measurement using AS-OCT is a simple and reproducible method that allows the direct comparison of numerical values. AS-OCT can be used to measure MRD-1 values for preoperative and postoperative values in patients with blepharoptosis. This measurement method might enable accurate and objective assessment of the degree of recovery and postoperative course following blepharoptosis surgery.

Disclosure

The author reports no conflicts of interest in this work.

References

1. Battu VK, Meyer DR, Wobig JL. Improvement in subjective visual function and quality of life outcome measures after blepharoptosis surgery. Am J Ophthalmol. 1996;121:677–686. doi:10.1016/S0002-9394(14)70634-8

2. Federici TJ, Meyer DR, Lininger LL. Correlation of the vision-related functional impairment associated with blepharoptosis and the impact of blepharoptosis surgery. Ophthalmology. 1999;106:1705–1712. doi:10.1016/S0161-6420(99)90354-8

3. Ho SF, Morawski A, Sampath R, et al. Modified visual field test for ptosis surgery (Leicester Peripheral Field Test). Eye. 2011;25:365–369. doi:10.1038/eye.2010.210

4. Cahill KV, Bradley EA, Meyer DR, et al. Functional indications for upper eyelid ptosis and blepharoplasty surgery: a report by the American Academy of Ophthalmology. Ophthalmology. 2011;118:2510–2517. doi:10.1016/j.ophtha.2011.09.029

5. Pemberton JD, Salter M, Fay A, et al. Investigation of goldmann perimetry in evaluation of patients for upper eyelid blepharoplasty. Orbit. 2018;37:48–52. doi:10.1080/01676830.2017.1353115

6. Boboridis K, Assi A, Indar A, et al. Repeatability and reproducibility of upper eyelid measurements. Br J Ophthalmol. 2001;85:99–101. doi:10.1136/bjo.85.1.99

7. Choi CJ, Lefebvre DR, Yoon MK. Validation of the facial assessment by computer evaluation (FACE) program for software-aided eyelid measurements. Orbit. 2016;35:117–120. doi:10.3109/01676830.2016.1139595

8. Volk Eye check―ocular measurement device. Available from: https://volk.com/media/wysiwyg/pdf/ifu/IM-078/IM-078_English.pdf. Accesed July 5, 2017.

9. Choi CJ, Chou JC, Lefebvre DR, et al. Margin reflex distance: differences based on camera and flash positions. Ophthalmic Plast Reconstr Surg. 2016;32:199–203. doi:10.1097/IOP.0000000000000456

10. Rüfer F, Schröder A, Erb C. White-to-white corneal diameter: normal values in healthy humans obtained with the Orbscan II topography system. Cornea. 2005;24:259–261. doi:10.1097/01.ico.0000148312.01805.53

11. Ogasawara K. Ptosis surgery with an improved method of shortening the Müller’s muscle and the levator aponeurosis with transconjunctival threading. J Ophthalmic Surg. 2019;2:7–12.

12. Tomey Corporation. Fourier domain OCT CASIA2. Available from: http://tomey.de/images/product_flyer/CASIA2_br_w.pdf. Accessed November 21, 2017.

13. Xu BY, Mai DD, Penteado RC, et al. Reproducibility and agreement of anterior segment parameter measurements obtained using the CASIA2 and spectralis OCT2 optical coherence tomography devices. J Glaucoma. 2017;26(11):974–979. doi:10.1097/IJG.0000000000000788

14. Zheng X, Goto T, Shiraishi A, et al. New method to analyze sagittal images of upper eyelid obtained by anterior segment optical coherence tomography. Orbit. 2019;38(6):446–452. doi:10.1080/01676830.2018.1563200

15. Rajyalakshmi R, Prakash WD, Ali MJ, et al. Periorbital biometric measurements using ImageJ software: standardisation of technique and assessment of intra- and interobserver variability. J Cutan Aesthet Surg. 2017;10:130–135. doi:10.4103/JCAS.JCAS_61_17

16. Sinha KR, Yeganeh A, Goldberg RA, et al. Assessing the accuracy of eyelid measurements utilizing the Volk eye check system and clinical measurements. Ophthalmic Plast Reconstr Surg. 2018;34:346–350. doi:10.1097/IOP.0000000000000991

Creative Commons License © 2020 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.