Validation of the integration of technology that measures additional “vascular” indices into an ambulatory blood pressure monitoring system

Authors Kotovskaya Y, Kobalava Z, Orlov A

Received 4 February 2014

Accepted for publication 16 March 2014

Published 2 May 2014 Volume 2014:7 Pages 91—97

DOI https://doi.org/10.2147/MDER.S61839

Checked for plagiarism Yes

Review by Single-blind

Peer reviewer comments 3


Yulia V Kotovskaya,1 Zhanna D Kobalava,1 Artemy V Orlov2

1Propedeutics Department, Peoples’ Friendship University of Russia, 2Competitive System Analysis Department (No 65), National Research Nuclear University MEPhI, Moscow, Russia

Background: The objective of this study was to validate the novel integration of oscillometric (Vasotens®) technology into a BPLab® ambulatory blood pressure (BP) monitoring system to measure central BP, the aortic augmentation index, and pulse wave velocity (PWV) compared with the recommended and widely accepted tonometric method.
Methods: The ARTERY Society guidelines for comparison of PWV measurement techniques were used as the basis for recruitment of 99 individuals (mean age 44±19 years, 52 males). The standard for comparison was the conventional “classic” SphygmoCor device.
Results: Accordance of the two methods was satisfactory (r=0.98, mean difference of 2.9±3.5 mmHg for central systolic BP; r=0.98, mean difference of −1.1±2.3 mmHg for central diastolic BP; r=0.83, mean difference of −2.6%±13% for aortic augmentation index; r=0.85, mean difference of 0.69±1.4 for PWV).
Conclusion: The performance of Vasotens algorithms using an oscillometric ambulatory BP monitoring system is feasible for accurate diagnosis, risk assessment, and evaluation of the effects of antihypertensive drugs.

Keywords: validation study, ambulatory, 24-hour, monitoring, arterial stiffness, pulse wave velocity, augmentation index, central blood pressure

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.