Back to Journals » Clinical Ophthalmology » Volume 14

Effectiveness and Safety of Long Duration versus Short Duration Diode Laser Transscleral Cyclophotocoagulation

Authors Chan JC, Chow SC, Lai JS

Received 27 August 2019

Accepted for publication 3 January 2020

Published 22 January 2020 Volume 2020:14 Pages 197—204

DOI https://doi.org/10.2147/OPTH.S228910

Checked for plagiarism Yes

Review by Single-blind

Peer reviewer comments 2

Editor who approved publication: Dr Scott Fraser


Jonathan Cheuk-hung Chan, Shing Chuen Chow, Jimmy Shiu-ming Lai

Department of Ophthalmology, University of Hong Kong, Hong Kong

Correspondence: Jonathan Cheuk-hung Chan
Department of Ophthalmology, LKS Faculty of Medicine, University of Hong Kong, Room 301, Block B, Cyberport 4, Hong Kong
Tel +852 3962 1405
Email jonnochan@yahoo.com

Purpose: To compare the efficacy and safety of diode laser transscleral cyclophotocoagulation using either the long duration or short duration protocol.
Methods: Retrospective series of 23 consecutive patients with glaucoma who underwent continuous-wave diode laser transscleral cyclophotocoagulation from August 2016 to July 2018 at a tertiary hospital in Hong Kong. Laser pulse duration for the long and short duration protocols was defined as 3.0– 4.0 and 1.5– 2.0 s, respectively.
Results: There were 15 male and 8 female Chinese subjects (23 eyes), age 49– 90 (71.3 ± 2.7), with 10 subjects that underwent long duration cyclophotocoagulation (power 1239.2 ± 78.3 mW, spots 13.9 ± 1.4) and 13 subjects that had short duration cyclophotocoagulation (mean power 1817.3 ± 85.7 mW, spots 14.4 ± 1.0). Six months after long and short duration cyclophotocoagulation, intraocular pressure decreased significantly from 29.9 ± 7.8 to 21.1 ± 6.5 (p < 0.01), and from 35.4 ± 2.7 to 24.1 ± 3.4 (p = 0.04), respectively, while glaucoma medications decreased significantly by 1.4 ± 0.5 (p = 0.02) in the long duration group only. Reduction of medications after short duration cyclophotocoagulation was less and did not reach statistical significance (0.9 ± 0.9, p = 0.15). There was no significant difference of visual deterioration and complication rates.
Conclusion: Both types of cyclophotocoagulation were equally effective in lowering intraocular pressure by 6 months, but the short duration protocol, using higher laser power, was able to achieve a greater and earlier reduction, at 3 months. However, the long duration protocol, using less laser power, appears better at reducing medication requirement by 6 months.

Keywords: cyclodiode, cycloablation, cyclodestruction, cyclodestructive, g-probe


Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.