Comparison of efficacy, safety, and predictability of laser in situ keratomileusis using two laser suites

Authors Meidani A, Tzavara C

Received 15 April 2016

Accepted for publication 21 June 2016

Published 24 August 2016 Volume 2016:10 Pages 1639—1646

DOI https://doi.org/10.2147/OPTH.S110626

Checked for plagiarism Yes

Review by Single-blind

Peer reviewers approved by Dr Colin Mak

Peer reviewer comments 4

Editor who approved publication: Dr Scott Fraser


Alexandra Meidani,1–3 Chara Tzavara3

1Hypervision Laser Centre, 2Eye Day Clinic, 3Department of Hygiene, University of Athens Medical School, Centre for Health Services Research, Epidemiology and Medical Statistics, Athens, Greece

Purpose: The main aim of this study was to compare the efficacy, safety, and predictability of femtosecond laser-assisted in situ keratomileusis performed by two different laser suites in the treatment of myopia for up to 6 months.
Methods: In this two-site retrospective nonrandomized study, myopic eyes that underwent laser-assisted in situ keratomileusis using IntraLase FS 60 kHz formed group 1 and those using WaveLight FS200 femtosecond laser system formed group 2. Ablation was performed with Visx Star S4 IR and WaveLight EX500 Excimer lasers, respectively, in groups 1 and 2. Both groups were well matched for age, sex, and mean level of preoperative refractive spherical equivalent (MRSE). Uncorrected distance visual acuity, corrected distance visual acuity, and MRSE were evaluated preoperatively and at 1 week, 1 month, and 6 months after treatment.
Results: Fifty-six eyes of 28 patients were included in the study. At 6-month follow-up postop, 78.6% of eyes in group 1 and 92.8% of eyes in group 2 achieved an uncorrected distance visual acuity of 20/20 or better (P=0.252). 35.7% and 50% in group 1 and group 2, respectively, gained one line (P=0.179). No eye lost lines of corrected distance visual acuity. Twenty-five eyes in group 1 (92.7%) and 27 eyes in group 2 (96.3%) had MRSE within ±0.5 D in the 6-month follow-up (P>0.999). The mean efficacy index at 6 months was similar in group 1 and group 2 (mean 1.10±0.12 [standard deviation] vs 1.10±0.1) (P=0.799). The mean safety index was similar in group 1 and group 2 (mean 1.10±0.10 [standard deviation] vs 1.10±0.09) (P=0.407).
Conclusion: The outcomes were excellent between the two laser suites. There were no significant differences at 6-month follow-up postop between the two laser systems.

Keywords: myopia, LASIK, IntraLase FS 60 kHz, WaveLight FS200, Visx Star S4 IR, WaveLight EX500
 

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.