Are adrenaline autoinjectors fit for purpose? A pilot study of the mechanical and injection performance characteristics of a cartridge- versus a syringe-based autoinjector
Original Research
(1525) Views (541) Full article downloads
Authors: Andreas Schwirtz, Harald Seeger
Published Date November 2010
Volume 2010:3 Pages 159 - 167
DOI: http://dx.doi.org/10.2147/JAA.S14419
Andreas Schwirtz, Harald SeegerPharma Consult GmbH, Vienna, Austria
Background: Adrenaline autoinjectors (AAIs) are prescribed to facilitate the intramuscular administration of adrenaline in patients diagnosed with life-threatening anaphylaxis. This pilot study investigated the injection and functional properties of two AAIs (deploying different delivery systems) under standard conditions, after dynamic and mechanical stresses, and in the presence of denim.
Methods: The differences between a cartridge-based AAI (EpiPen® Junior) and a syringe-based AAI (Anapen® Junior) were assessed using three sets of tests. Test 1: under standard conditions, the injection depth and dose were measured in ballistic gelatine (a validated tissue simulant). Test 2: before the safety cap removal and activation forces were measured, AAIs were subjected to either of two preconditioning tests: 1) free-fall drop test; or 2) static load (ie, 400 N, equivalent to 40 kg weight) test; or 3) no preconditioning. Test 3: under standard conditions, injection properties into ballistic gelatine in the presence and absence of denim were investigated. Statistical analyses were performed using the Student’s t-test or Welch’s test.
Results: The maximum depth of delivery was significantly greater with cartridge AAI (n = 4, mean 21.09 ± 2.54 mm) than with syringe AAI (n = 5; mean 11.64 ± 0.80 mm; P = 0.003). After 2.5 seconds, cartridge AAI (n = 4) discharged significantly more dose than syringe AAI (n = 3; 74.3% versus 25.7% of total dose; P = 0.001). Both cartridge and syringe AAI withstood the free-fall drop test, but almost all devices failed to activate following the static load test. Under standard conditions, significantly less force was required to remove the safety cap of cartridge AAI than syringe AAI (both n = 15; mean 9.56 ± 2.36 N versus 20.23 ± 6.61 N, respectively; P < 0.001), but a significantly greater activation force was required for cartridge AAI than syringe AAI (mean 23.01 ± 3.96 N versus 8.06 ± 0.51 N, respectively; P < 0.001). The presence of denim did not alter the activation force or effective needle length of either of the AAIs.
Conclusion: Cartridge AAI appears significantly more capable of consistently and rapidly delivering a clinically relevant dose of intramuscular adrenaline than syringe AAI. However, both devices showed shortcomings in their ability to sustain mechanical stress similar to that which is likely over their shelf life, and as such, may not be fit for life-saving purpose.
Keywords: anaphylaxis, intramuscular, delivery system, device
Readers of this article also read:
Updates in the treatment of ocular allergies
Current recommendations for the treatment of mild asthma
New therapies and management strategies in the treatment of asthma: patient-focused developments
The current role of sublingual immunotherapy in the treatment of allergic rhinitis in adults and children
Effectiveness of inhaler types for real-world asthma management: retrospective observational study using the GPRD
Comparison of stereotactic core breast biopsy and open surgical biopsy results at a tertiary care hospital in Pakistan
Childhood asthma and anthropogenic CO2 emissions
Exploring UK attitudes towards unlicensed medicines use: a questionnaire-based study of members of the general public and physicians
Management of critically ill patients receiving noninvasive and invasive mechanical ventilation in the emergency department
- Testimonials
"You do a tremendous job!!" Ruben Restrepo, The University of Texas Health Science Center at San Antonio
- Clinical effectiveness of the Respimat® inhaler device in managing chronic obstructive pulmonary disease: evidence when compared with other handheld inhaler devices
- Chronic obstructive pulmonary disease as an independent risk factor for cardiovascular morbidity
- The pathophysiology of bronchiectasis
- Exacerbation rate, health status and mortality in COPD – a review of potential interventions