Optimizing visualization in enhanced depth imaging OCT in healthy subjects and patients with retinal pigment epithelial detachment
Authors Reznicek L, Vounotrypidis, Seidensticker F, Kortüm K , Kampik A, Neubauer A, Wolf A
Received 4 July 2012
Accepted for publication 11 August 2012
Published 21 November 2012 Volume 2012:6 Pages 1915—1920
DOI https://doi.org/10.2147/OPTH.S35596
Checked for plagiarism Yes
Review by Single anonymous peer review
Peer reviewer comments 2
Lukas Reznicek, Efstathios Vounotrypidis, Florian Seidensticker, Karsten Kortuem, Anselm Kampik, Aljoscha S Neubauer, Armin Wolf
Department of Ophthalmology, Ludwig Maximilians University Muenchen, Munich, Germany
Background: This study’s objective was to optimize the visualization of three different spectral-domain optical coherence tomography (SD-OCT) display modalities and evaluate enhanced depth imaging (EDI) by comparing the maximum depth of assessment in conventional versus inverted cross-sectional OCT images in healthy subjects and in patients with retinal pigment epithelial detachment (PED).
Methods: Cross-sectional SD-OCT conventional and inverted images were obtained with the HRA2 (Heidelberg Retina Angiograph II, Heidelberg Engineering, Heidelberg, Germany). Horizontal as well as vertical sections in three different display modes were blinded for evaluation by three independent, experienced graders for maximal imaging depth of the deep ocular fundus layers.
Results: The mean imaging depth as measured from the inner segment/outer segment (IS/OS) to the outer choroid of all 14 healthy subjects was 197 ± 44 µm vs 263 ± 56 µm for conventional vs EDI scans: in black/white mode, it was significantly lower (P < 0.001) than in white/black mode (249 ± 42 µm vs 337 ± 71 µm) and color/heat mode (254 ± 48 µm vs 354 ± 73 µm). The mean imaging depth of all 14 study eyes with PED was 240 ± 78 µm vs 345 ± 100 µm for conventional vs EDI scans in black/white mode, and was significantly lower (P < 0.001) than in white/black mode (393 ± 104 µm vs 464 ± 126 µm) and in color/heat mode (373 ± 106 µm vs 453 ± 114 µm). In each display modality of healthy subjects and of patients with PED, EDI scans showed a significantly higher imaging depth than the corresponding conventional scans.
Conclusion: White/black and color/heat modes allow increased imaging depth, compared to black/white mode using both conventional or EDI OCT scans in healthy subjects or patients with PED. EDI obtained with HRA2 significantly improves the imaging depth, compared to conventional OCT scans.
Keywords: OCT, EDI OCT, enhanced depth imaging, pigment epithelial detachment, imaging mode, optical coherence tomography
© 2012 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.