Variability of standardized echographic ultrasound using 10 mHz and high-resolution 20 mHz B scan in measuring intraocular melanoma

Authors Kook D, Kreutzer T, Wolf A, Haritoglou C

Published 18 April 2011 Volume 2011:5 Pages 477—482

DOI https://doi.org/10.2147/OPTH.S18513

Review by Single anonymous peer review

Peer reviewer comments 2



Daniel Kook*, Thomas C Kreutzer*, Armin Wolf, Christos Haritoglou
Department of Ophthalmology, Ludwig-Maximilians University, Munich, Germany
*These authors contributed equally

Background: The purpose of this study was to evaluate the applicability and variability of echographic imaging using 10 mHz and high-resolution 20 mHz B scanning for measurement of intraocular tumors.
Methods: This prospective consecutive study comprises 27 eyes in 27 patients with uveal melanoma. Each patient was examined by three experienced examiners independently on three occasions within a two-week time frame in a blinded manner. The height of the lesion was measured by all examiners using the 10 mHz B, 20 mHz B, and 8 mHz A scan probes. Additionally, basal diameter was examined using the 10 and 20 mHz B scan.
Results: Tumor height measurements for all examiners using the standardized A scan tended to be higher than for both B scan measurements. Statistical analysis revealed significant differences in tumor height between B and A scan measurements. No difference in tumor height was found between the two B scan techniques (P = 0.239). Basal tumor diameter measurements revealed significant differences between 10 mHz and 20 mHz B scans (P < 0.001 and P = 0.001, respectively). For the 10 mHz B scan, basal diameter results tended to be larger than for the 20 mHz B scan. No difference was found for interobserver variation in all A scan and B scan examinations. The mean standard deviation of the difference in tumor height measurements between the examiners was ±0.24 mm for the 8 mHz A scan, ±0.46 mm for the 10 mHz B scan, and ±0.42 mm for the 20 mHz B scan. Both the 10 mHz and more precise 20 mHz B scan evaluations underestimated tumor height.
Conclusion: The 20 mHz ultrasound probe, despite its theoretically higher resolution, is not able to replace A scan measurements of tumor height.

Keywords: standardized echography, 20 mHz, B scan, A scan, intraocular tumor

Creative Commons License © 2011 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.