Validation of four devices: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 for home blood pressure measurement according to the European Society of Hypertension International Protocol

Authors Topouchian J, Agnoletti D, Blacher J, Youssef A, Chahine MN, Ibanez I, Assemani N, Asmar R

Received 4 September 2013

Accepted for publication 9 November 2013

Published 16 January 2014 Volume 2014:10 Pages 33—44

DOI https://doi.org/10.2147/VHRM.S53968

Checked for plagiarism Yes

Review by Single-blind

Peer reviewer comments 10

Video abstract presented by Mirna Chahine

Views: 608

Jirar Topouchian,1 Davide Agnoletti,1 Jacques Blacher,1 Ahmed Youssef,1 Mirna N Chahine,2,3 Isabel Ibanez,3 Nathalie Assemani,3 Roland Asmar1–3

1
Centre de Diagnostic, Hôpital Hôtel-Dieu, Paris, France; 2Faculty of Medicine of the Lebanese University, 3Foundation-Medical Research Institutes, Beirut, Lebanon

Background: Four oscillometric devices, including the Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670, designed for self-blood pressure measurement (SBPM) were evaluated according to the European Society of Hypertension (ESH) International Protocol Revision 2010 in four separate studies.
Methods: The four devices measure brachial blood pressure (BP) using the oscillometric method. The Withings BP-800 has to be connected to an Apple® iOS device such as an iPhone®, iPad®, or iPod®. The ESH International Protocol Revision 2010 includes a total number of 33 subjects. The difference between observer and device BP values was calculated for each measure. Ninety-nine pairs of BP differences were classified into three categories (≤5 mmHg, ≤10 mmHg, ≤15 mmHg). The protocol procedures were followed precisely in each of the four studies.
Results: All four tested devices passed the validation process. The mean differences between the device and mercury readings were: −1.8±5.1 mmHg and −0.4±2.8 mmHg for systolic and diastolic BP, respectively, using the Omron M6 Comfort device; 2.5±4.6 mmHg and −1.2±4.3 mmHg for the Omron HEM-7420 device; −0.2±5.0 mmHg and 0.4±4.2 mmHg for the Withings BP-800 device; and 3.0±5.3 mmHg and 0.3±5.2 mmHg for the Polygreen KP-7670 device.
Conclusion: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 readings differing by less than 5 mmHg, 10 mmHg, and 15 mmHg fulfill the ESH International Protocol Revision 2010 requirements, and therefore are suitable for use by patients for SBPM, if used correctly.

Keywords: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, Polygreen KP-7670, self-blood pressure measurement, validation, European Society of Hypertension International Protocol Revision

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.

Download Article [PDF]  View Full Text [HTML][Machine readable]

 

Readers of this article also read:

Increased serum levels of interleukin-10 predict poor prognosis in extranodal natural killer/T-cell lymphoma patients receiving asparaginase-based chemotherapy

Wang H, Wang L, Wuxiao ZJ, Huang HQ, Jiang WQ, Li ZM, Lu Y, Xia ZJ

OncoTargets and Therapy 2015, 8:2589-2599

Published Date: 14 September 2015

Companion diagnostics and molecular imaging-enhanced approaches for oncology clinical trials

Van Heertum RL, Scarimbolo R, Ford R, Berdougo E, O’Neal M

Drug Design, Development and Therapy 2015, 9:5215-5223

Published Date: 11 September 2015

Ageism and its clinical impact in oncogeriatry: state of knowledge and therapeutic leads

Schroyen S, Adam S, Jerusalem G, Missotten P

Clinical Interventions in Aging 2015, 10:117-125

Published Date: 31 December 2014

Green synthesis of water-soluble nontoxic polymeric nanocomposites containing silver nanoparticles

Prozorova GF, Pozdnyakov AS, Kuznetsova NP, Korzhova SA, Emel’yanov AI, Ermakova TG, Fadeeva TV, Sosedova LM

International Journal of Nanomedicine 2014, 9:1883-1889

Published Date: 16 April 2014

Cross-linked acrylic hydrogel for the controlled delivery of hydrophobic drugs in cancer therapy

Deepa G, Thulasidasan AK, Anto RJ, Pillai JJ, Kumar GS

International Journal of Nanomedicine 2012, 7:4077-4088

Published Date: 27 July 2012

Pegylated liposomal doxorubicin in the management of ovarian cancer

Gabriella Ferrandina, Giacomo Corrado, Angelo Licameli, et al

Therapeutics and Clinical Risk Management 2010, 6:463-483

Published Date: 29 September 2010

Crystallization after intravitreal ganciclovir injection

Pitipol Choopong, Nattaporn Tesavibul, Nattawut Rodanant

Clinical Ophthalmology 2010, 4:709-711

Published Date: 14 July 2010