Validation and verification of the OPI 2.0 System

Authors Abelson R, Lane K, Rodriguez J , Johnston P, Angjeli E, Ousler, Montgomery D

Received 23 December 2011

Accepted for publication 21 January 2012

Published 24 April 2012 Volume 2012:6 Pages 613—622

DOI https://doi.org/10.2147/OPTH.S29431

Review by Single anonymous peer review

Peer reviewer comments 3



Richard Abelson1,2, Keith J Lane3, John Rodriguez3, Patrick Johnston3, Endri Angjeli3, George Ousler3, Douglas Montgomery1
1School of Computing, Informatics, and Decision Systems Engineering, Arizona State University, Tempe, AZ, 2Statistics and Data Corporation, Tempe, AZ, 3Ora, Inc, Andover, MA, USA

Purpose: The Ocular Protection Index (OPI) 2.0 System was developed to evaluate ocular surface protection under a natural blink pattern and normal visual conditions. The OPI 2.0 System implements fully automated software algorithms which provide a real-time measurement of corneal exposure (breakup area) for each interblink interval during a 1-minute video. Utilizing this method, the mean breakup area (MBA) and OPI 2.0 (MBA/interblink interval) were calculated and analyzed. The purpose of this study was to verify and validate the OPI 2.0 System for its ability to distinguish between dry eye and normal subjects, and to accurately identify breakup area.
Methods: In order to verify and validate the OPI 2.0 System, a series of artificial images and a series of still image frames captured during an actual clinical session using fluorescein staining videography were analyzed. Finally, a clinical validation process was completed to determine the effectiveness and clinical relevance of the OPI 2.0 System to differentiate between dry eye and normal subjects.
Results: Software analysis verification conducted in a set of artificially constructed images and in actual videos both saw minimal error rates. MBA and OPI 2.0 calculations were able to distinguish between the qualifying eyes of dry eye and normal subjects in a statistically significant fashion (P < 0.001 for both outcomes). As expected, dry eye subjects had a higher MBA and OPI 2.0 than normal subjects (0.232, dry eye; 0.040, normal and 0.039, dry eye; 0.006, normal, respectively). Results for the worst eyes and all qualifying analyses based on staining, forced-stare tear film breakup time, and MBA were numerically similar.
Conclusion: The OPI 2.0 System accurately identifies the degree of breakup area on the cornea and represents an efficient, clinically relevant measurement of the pathophysiology of the ocular surface.

Keywords: tear film breakup time, interblink interval, natural blink conditions, Ocular Protection Index

Creative Commons License © 2012 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.