The use of Invos™ somatic oximetry to measure variations in placental tissue oxygenation in laboring healthy term parturients with epidural analgesia: an observational study

Authors Loubert C, Ouellette M, Zaphiratos V, Tanoubi I

Received 31 August 2017

Accepted for publication 19 September 2017

Published 25 October 2017 Volume 2017:10 Pages 253—256

DOI https://doi.org/10.2147/MDER.S150472

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 2

Editor who approved publication: Dr Scott Fraser


Christian Loubert,1 Mélissa Ouellette,2 Valérie Zaphiratos,1 Issam Tanoubi1

1Department of Anesthesia, Maisonneuve-Rosemont Hospital, CIUSSS-de-l’Est de l’île de Montréal, University of Montreal, Montreal, 2Department of Anesthesia, Saint-Jerôme Hospital Center, CISSS des Laurentides, Saint-Jerôme, QC, Canada

Purpose:
Near-infrared spectroscopy provides a non-invasive continuous real-time monitoring of tissue oxygen saturation. As uterine contractions during labor may be associated with a transient uteroplacental hypoperfusion, this prospective, observational study investigates the ability of near-infrared spectroscopy to detect variation in uteroplacental oximetry during uterine contractions.
Patients and methods: Four Invos™ oximetry probes (Medtronic®, Minneapolis, MN, USA) per subjects were applied on the placental surface (PLA), the abdomen (MYO), the forearm (ARM) and the leg (LEG), of twenty healthy laboring parturients with epidural analgesia. Measurements of mean tissue oxygen saturation and area under the curve (AUC) were made during 60 minutes. The primary outcome was the difference of the AUC measurements between the PLA probe and the MYO probe.
Results: The AUC values for the PLA and MYO probes were not different. The mean saturation values recorded by the PLA probe were not different from the other probes.
Conclusion: The Invos monitor was unable to detect variations in uteroplacental saturation during labor in healthy parturients.

Keywords: near-infrared spectroscopy, uteroplacental perfusion, Invos™ monitor, tissue oxygenation

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.