The threshold elicited motor response via an in situ femoral nerve catheter predicts analgesia following total knee replacement

Authors Nader A, Kendall M, Chung B, Chekka K, Malik K, McCarthy RJ

Published 14 September 2010 Volume 2010:3 Pages 109—114

DOI https://doi.org/10.2147/LRA.S12208

Review by Single anonymous peer review

Peer reviewer comments 2



Antoun Nader, Mark C Kendall, Brian Chung, Kiran Chekka, Khalid Malik, Robert J McCarthy
Department of Anesthesiology, Northwestern University Feinberg School of Medicine, Chicago, IL, USA

Background: Continuous femoral nerve blocks are a part of the multimodal perioperative anesthetic regimen following total knee replacement. Elicitation of a quadriceps muscle contraction (QC) at placement is desirable. We prospectively evaluated the relationship between elicited motor response and threshold current with block success in situ femoral nerve catheters after total knee replacement.
Methods: After Institutional Review Board approval and written informed consent, 100 adults aged ≥18 years, undergoing total knee replacement were studied. The threshold current for an elicited motor response (QC or sartorius muscle contraction [SC]) was recorded during needle insertion, after femoral nerve catheter advancement, and prior to local anesthetic administration on the first postoperative day. Patients were assessed for pinprick sensory anesthesia of the femoral nerve distribution by an observer unaware of the current threshold or evoked motor response at 5-minute intervals for 30 minutes. A successful block was considered a lack of pinprick sensation within 30 minutes.
Results: Ninety patients completed the study. A QC was obtained at needle insertion in 89 and maintained following catheter advancement in 77 subjects. Prior to drug administration a QC was obtained in 66 patients, 13 demonstrated an SC, and 11 had no motor response. QC prior to drug administration had a sensitivity of 0.85 (95% confidence intervals [CI]: 0.75 to 0.92) and a specificity of 0.55 (95% CI: 0.37 to 0.72) for complete femoral nerve block. The specificity of a QC prior to drug administration was greater than at catheter placement (P = 0.006). Compared to the threshold current at catheter placement, 27 patients had a decrease, 60 had an increase, and 3 exhibited no change in the threshold current prior to drug administration (P < 0.01). The median interquartile range (IQR) threshold current in patients that achieved a complete block was 0.56 (0.29 to 0.80) mA compared with 1.1 mA (0.41 to 2.75) mA for incomplete blocks (P < 0.01). The area under the receiver operator characteristics curve for current threshold prior to local anesthetic administration (0.74) was greater than at catheter placement (0.45) (P < 0.001). The intersection of sensitivity and specificity for the minimal threshold current prior to local anesthetic injection was 0.84 mA.
Conclusion: The elicited motor response and current threshold from a stimulating femoral catheter measured prior to local anesthetic injection is an important determinant of the success of femoral nerve block following bolus administration.

Keywords: stimulating femoral nerve catheters, total knee replacement, postoperative analgesia

Creative Commons License © 2010 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.