Safety and effectiveness of a glistening-free single-piece hydrophobic acrylic intraocular lens (enVista)

Authors Packer M , Fry L, Lavery KT, Lehmann R , McDonald J, Nichamin L, Bearie B, Hayashida J, Altmann GE, Khodai O

Received 25 June 2013

Accepted for publication 24 July 2013

Published 24 September 2013 Volume 2013:7 Pages 1905—1912

DOI https://doi.org/10.2147/OPTH.S50499

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 3



Mark Packer,1 Luther Fry,2 Kevin T Lavery,3 Robert Lehmann,4 James McDonald,5 Louis Nichamin,6 Brian Bearie,7,† Jon Hayashida,8 Griffith E Altmann,8 Omid Khodai8

1Department of Ophthalmology, Oregon Health and Science University, Eugene, OR, USA; 2University of Kansas Medical Center, Kansas City, KS, USA; 3Wayne State University, Detroit, MI, USA; 4Baylor College of Medicine, Houston, TX, USA; 5University of Arkansas for Medical Sciences, Little Rock, AR, USA; 6Laurel Eye Clinic, Brookville, PA, USA; 7Grand Rapids Eye Institute, Grand Rapids, MI, USA; 8Bausch & Lomb, Aliso Viejo, CA, USA

Brian Bearie passed away on March 9, 2011

Purpose: To evaluate the safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (IOL; enVista model MX60; Bausch & Lomb, Rochester, NY, USA) when used to correct aphakia following cataract extraction in adults.
Methods: This was a prospective case series (NCT01230060) conducted in private practices in the US. Eligible subjects were adult patients with age-related cataract amenable to treatment with standard phacoemulsification/extracapsular cataract extraction. With follow-up of 6 months, primary safety and effectiveness end points included the rates of US Food and Drug Administration (FDA)-defined cumulative and persistent adverse events and the percentage of subjects who achieved best-corrected visual acuity (BCVA) of 20/40 or better at final visit. To evaluate rotational stability, subjects were randomized (1:1:1:1) to have the lens implanted in one of four axis positions in 45° increments.
Results: A total of 122 subjects were enrolled. The rate of cumulative and persistent adverse events did not significantly exceed historical controls, as per FDA draft guidance. At the final postoperative visit, all subjects (100%) achieved a BCVA of 20/40 compared with the FDA historical control of 96.7%. Rotation of the IOL between the two final follow-up visits was ≤5° for 100% of eyes, and refractive stability was demonstrated. A low evaluation of posterior capsule opacification score was demonstrated, and no glistenings of any grade were reported for any subject at any visit.
Conclusion: This study demonstrated the safety and effectiveness of the MX60 IOL. Favorable clinical outcomes included preserved BCVA, excellent rotational and refractive stability, no glistenings, and a low evaluation of posterior capsule opacification score.

Keywords: cataract surgery, enVista, glistenings, intraocular lens, hydrophobic acrylic

Creative Commons License © 2013 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.