Refractive and Visual Outcomes After Bilateral Implantation of a Trifocal Intraocular Lens in a Large Population

Authors Carreño E, Carreño EA, Carreño R, Carreño M, López V, Potvin R

Received 16 November 2019

Accepted for publication 30 January 2020

Published 7 February 2020 Volume 2020:14 Pages 369—376

DOI https://doi.org/10.2147/OPTH.S238841

Checked for plagiarism Yes

Review by Single-blind

Peer reviewer comments 3

Editor who approved publication: Dr Scott Fraser


Edgardo Carreño,1 Edgardo A Carreño,1 Rodrigo Carreño,1 Maricarmen Carreño,1 Veronica López,1 Richard Potvin2

1Centro Oftalmológico Carreño, Las Condes, Santiago, Chile; 2Science in Vision, Akron, NY, USA

Correspondence: Richard Potvin
Science in Vision, 6197 Dye Road, Akron, NY 14001, USA
Email rick@scienceinvision.com

Purpose: To review refractive, visual acuity, defocus curve and contrast sensitivity results after bilateral implantation of a trifocal intraocular lens (IOL) in a large population.
Setting: One site in Santiago, Chile.
Design: Single arm, non-randomized retrospective chart review.
Methods: This was a single-arm retrospective chart review of clinical outcomes after bilateral implantation of a trifocal IOL (Panoptix®), both toric and non-toric versions. Binocular visual acuity at 4 m, 60 cm and 40 cm was tested. Other tests included refraction, mesopic and photopic contrast sensitivity, and defocus curve measurement.
Results: The review included 500 eyes of 250 patients implanted with the trifocal IOL and 200 eyes of 100 patients implanted with the trifocal toric IOL, with no clinically significant differences between groups. Ninety-six percent of all eyes were within 0.50D of the intended spherical equivalent correction. In the toric group, 94% of eyes (187/200) had a residual refractive cylinder ≤ 0.50D, compared to 81% of eyes (406/500) in the non-toric group. Four out of five patients (80.6%, 282/350) had a binocular uncorrected VA of 0.1 logMAR (20/25) at all test distances. Mean defocus was 0.1 logMAR or better from vergences from 0.00 to − 3.00 D (corresponding to vision from distance to about 33 cm). With a cutoff of 0.2 logMAR, 96% of patients had a range of vision 2.5 D or greater. Contrast sensitivity was similar between the toric and non-toric lenses, and similar to age-matched normal results.
Conclusion: The non-toric and toric trifocal IOLs provided good distance, intermediate and near vision to patients, with a wide range of vision and good contrast sensitivity.

Keywords: trifocal IOL, contrast sensitivity, toric trifocal IOL

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.