Back to Journals » Clinical Ophthalmology » Volume 6

Quantitative evaluation of corneal epithelial injury caused by n-heptanol using a corneal resistance measuring device in vivo

Authors Fukuda M, Sasaki

Received 16 February 2012

Accepted for publication 7 March 2012

Published 17 April 2012 Volume 2012:6 Pages 585—593

DOI https://doi.org/10.2147/OPTH.S30935

Review by Single anonymous peer review

Peer reviewer comments 3



Masamichi Fukuda1, Hiroshi Sasaki1
1Department of Ophthalmology, Kanazawa Medical University, Uchinada, Japan

Purpose: We attempted to develop a device for measuring electrical corneal resistance (CR) using corneal contact lens electrodes to quantitatively evaluate corneal injury in vivo. In the present study, full-thickness detachment of the corneal epithelium was induced by n-heptanol, and the feasibility of the quantitative evaluation of this injury by corneal contact lens electrodes was evaluated in vivo.
Methods: The central area of an albino rabbit's cornea was exposed to a filter paper pre-immersed in n-heptanol for 1 minute to induce injury of the corneal epithelium. After induction of injury, the electrical CR was measured and the percentage of CR (%CR) was calculated. Fluorescein solution (3 µL) was applied to the wound/affected area of the corneal epithelium for photography with a slit-lamp biomicroscope. The wound/affected area was measured using an image analysis system. The correlation between the %CR and the wound/affected area was analyzed.
Results: As the size of the wound/affected area of the corneal epithelium increased, the %CR decreased after corneal epithelium detachment. Thus, a close correlation was found between the area of corneal epithelium detachment and the %CR.
Conclusion: The corneal resistance device that we developed was capable of quantitatively evaluating n-heptanol-induced full-thickness injuries of the corneal epithelium.

Keywords: eyes in vivo, corneal injury, corneal contact lens electrode corneal resistance device, n-heptanol

Creative Commons License © 2012 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.