Pulse wave velocity 24-hour monitoring with one-site measurements by oscillometry

Authors Posokhov I

Received 25 December 2012

Accepted for publication 19 January 2013

Published 19 February 2013 Volume 2013:6 Pages 11—15

DOI https://doi.org/10.2147/MDER.S42082

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 2



Igor N Posokhov

Hemodynamic Laboratory Ltd, Nizhniy Novgorod, Russia


Abstract: This review describes issues for the estimation of pulse wave velocity (PWV) under ambulatory conditions using oscillometric systems. The difference between the principles of measuring the PWV by the standard method and by oscillometry is shown, and information on device validation studies is summarized. It was concluded that currently oscillometry is a method that is very convenient to use in the 24-hour monitoring of the PWV, is relatively accurate, and is reasonably comfortable for the patient. Several indices with the same principles as those in the analysis of blood pressure in ambulatory monitoring of blood pressure, namely the assessment of load, variability, and circadian rhythm, are proposed.

Keywords: pulse wave velocity, 24-hour monitoring, oscillometry

Creative Commons License © 2013 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.