New stent design for use in small coronary arteries during percutaneous coronary intervention

Authors Granada J, Huibregtse BA, Dawkins KD

Published 19 October 2010 Volume 2010:3 Pages 57—66

DOI https://doi.org/10.2147/MDER.S13494

Review by Single anonymous peer review

Peer reviewer comments 2



Juan F Granada1, Barbara A Huibregtse2, Keith D Dawkins2
1The Jack H Skirball Center for Cardiovascular Research, Cardiovascular Research Foundation, Columbia University Medical Center, New York, NY, USA; 2Boston Scientific Corporation, Natick, MA, USA

Abstract: Patients with diabetes mellitus, of female gender, increased age, and/or with peripheral vascular disease often develop coronary stenoses in small caliber vessels. This review describes treatment of these lesions with the paclitaxel-eluting 2.25 mm TAXUS® Liberté® Atom™ stent. Given the same stent composition, polymer, antirestenotic drug (paclitaxel), and release kinetics as the first-generation 2.25 mm TAXUS® Express® Atom™ stent, the second-generation TAXUS Liberté Atom stent incorporates improved stent design characteristics, including thinner struts (0.0038 versus 0.0052 inches), intended to increase conformability and deliverability. In a porcine noninjured coronary artery model, TAXUS Liberté Atom stent implantation in small vessels demonstrated complete strut tissue coverage compared with the bare metal stent control, suggesting a similar degree of tissue healing between the groups at 30, 90, and 180 days. The prospective, single-armed TAXUS ATLAS Small Vessel trial demonstrated improved instent late loss (0.28 ± 0.45 versus 0.84 ± 0.57 mm, P < 0.001), instent binary restenosis (13.0% versus 38.1%, P < 0.001), and target lesion revascularization (5.8% versus 17.6%, P < 0.001) at nine months with the TAXUS Liberté Atom stent as compared with the bare metal Express stent control, with similar safety measures between the two groups. The TAXUS Liberté Atom also significantly reduced nine-month angiographic rates of both instent late loss (0.28 ± 0.45 versus 0.44 ± 0.61 mm, P = 0.03) and instent binary restenosis (13.0% versus 25.9%, P = 0.02) when compared with the 2.25 mm TAXUS Express Atom control. The observed reduction in target lesion revascularization with the TAXUS Liberté Atom compared with the TAXUS Express Atom at nine months (5.8% versus 13.7%, P = 0.02) was sustained through three years (10.0% versus 22.1%, P = 0.008) with similar, stable safety outcomes between the groups. In conclusion, these data confirm the safety and favorable performance of the TAXUS Liberté Atom stent in the treatment of small coronary vessels.

Keywords: small vessel, paclitaxel, stent

Creative Commons License © 2010 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.