Back to Journals » Clinical Ophthalmology » Volume 4

Measuring accurate IOPs: Does correction factor help or hurt?

Authors Pinakin Gunvant, Robert D Newcomb, Elliot M Kirstein, et al

Published 28 June 2010 Volume 2010:4 Pages 611—616

DOI https://doi.org/10.2147/OPTH.S11105

Review by Single-blind

Peer reviewer comments 3

Pinakin Gunvant1, Robert D Newcomb2, Elliot M Kirstein3, Victor E Malinovsky4, Richard J Madonna5, Richard E Meetz 4

1Southern College of Optometry, Memphis, Tennessee, USA; 2The Ohio State University College of Optometry, Columbus, Ohio, USA; 3The Glaucoma and Diabetes Eye Institute, Cincinnati, Ohio, USA; 4Indiana University School of Optometry, Bloomington, Indiana, USA; 5State University of New York College of Optometry, New York, USA

Purpose: To evaluate if using the Ehlers correction factor on the intraocular pressure (IOP) measured using the Goldmann applanation tonometer (GAT) improves its agreement with the PASCAL dynamic contour tonometer (DCT).

Patients and methods: A total of 120 eyes of 120 individuals were examined. Participants underwent IOP measurement with both the DCT and the GAT and central corneal thickness measurement. The Ehlers correction factor was applied on the GAT IOP measurements to calculate Ehlers-corrected GAT IOP. The agreement between the DCT and GAT, and DCT and Ehlers-corrected GAT IOP was analyzed. The analyses were repeated by stratifying the data by race.

Results: The mean IOP of the GAT, DCT, and the Ehlers-corrected GAT was 15.30, 16.78, and 14.68 mmHg, respectively. The agreement as assessed by Bland–Altman plot for the GAT with the DCT and DCT and Ehlers-corrected GAT IOP was +4.1 to −6.9 and +4.15 to −8.25 mmHg, respectively. The results were similar even when stratifying the data by race.

Conclusion: Using Ehlers correction factor to account for the effect of corneal parameters on the IOP measured by the GAT worsens the agreement with the DCT. This effect remains even when stratifying the data by race.

Keywords: dynamic contour tonometer, Goldmann applanation tonometer, tonometric correction factors, central corneal thickness, intraocular pressure

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.

Download Article [PDF] 

 

Other articles by this author:

Macular pigment optical density: repeatability, intereye correlation, and effect of ocular dominance

Davey PG, Alvarez SD, Lee JY

Clinical Ophthalmology 2016, 10:1671-1678

Published Date: 29 August 2016

Fabry disease: a survey of visual and ocular symptoms

Davey PG

Clinical Ophthalmology 2014, 8:1555-1560

Published Date: 19 August 2014

Predicting the necessity of LASIK enhancement after cataract surgery in patients with multifocal IOL implantation

Gunvant P, Ablamowicz A, Gollamudi S

Clinical Ophthalmology 2011, 5:1281-1285

Published Date: 8 September 2011

Readers of this article also read:

Green synthesis of water-soluble nontoxic polymeric nanocomposites containing silver nanoparticles

Prozorova GF, Pozdnyakov AS, Kuznetsova NP, Korzhova SA, Emel’yanov AI, Ermakova TG, Fadeeva TV, Sosedova LM

International Journal of Nanomedicine 2014, 9:1883-1889

Published Date: 16 April 2014

Methacrylic-based nanogels for the pH-sensitive delivery of 5-Fluorouracil in the colon

Ashwanikumar N, Kumar NA, Nair SA, Kumar GS

International Journal of Nanomedicine 2012, 7:5769-5779

Published Date: 15 November 2012

A novel preparation method for silicone oil nanoemulsions and its application for coating hair with silicone

Hu Z, Liao M, Chen Y, Cai Y, Meng L, Liu Y, Lv N, Liu Z, Yuan W

International Journal of Nanomedicine 2012, 7:5719-5724

Published Date: 12 November 2012

Cross-linked acrylic hydrogel for the controlled delivery of hydrophobic drugs in cancer therapy

Deepa G, Thulasidasan AK, Anto RJ, Pillai JJ, Kumar GS

International Journal of Nanomedicine 2012, 7:4077-4088

Published Date: 27 July 2012

Crystallization after intravitreal ganciclovir injection

Pitipol Choopong, Nattaporn Tesavibul, Nattawut Rodanant

Clinical Ophthalmology 2010, 4:709-711

Published Date: 14 July 2010