Laboratory assessment of thermal characteristics of three phacoemulsification tip designs operated using torsional ultrasound

Authors Zacharias J

Received 26 January 2016

Accepted for publication 31 March 2016

Published 13 June 2016 Volume 2016:10 Pages 1095—1101

DOI https://doi.org/10.2147/OPTH.S105065

Checked for plagiarism Yes

Review by Single-blind

Peer reviewers approved by Dr Pelin Ozyol

Peer reviewer comments 3

Editor who approved publication: Dr Scott Fraser


Jaime Zacharias

Phacodynamics Laboratory, Pasteur Ophthalmic Clinic, Santiago, Chile

Purpose: Ultrasound activation of phacoemulsification (phaco) tips can create considerable thermal energy that may increase the risk of tissue damage during cataract surgery. The purpose of this study was to define the thermal profiles of three phaco tip designs in simulated surgical conditions.
Methods: In this laboratory investigation, sleeved phaco tips (mini-flared Kelman® tip with aspiration bypass port and Intrepid® Balanced Tip with aspiration bypass port, and MST A1 bent-mini phaco tip (without aspiration bypass) were tested using an ultrasonic phaco device operated in torsional mode at power levels of 50%, 75%, and 100% amplitude. An automated fixture applied a 30 g load to simulate compression against the incision site, leading to friction between the silicone sleeve and the titanium tip. Temperature was recorded by high rate infrared imaging under conditions of free flow and occlusion, which was simulated by clamping the aspiration line. Data were summarized using descriptive statistics.
Results: Baseline temperatures of ~26°C were observed for all tips. During ultrasonic operation at 50%, 75%, and 100% amplitude, temperatures were lower for the mini-flared and balanced tips versus the bent-mini tip, both when load was applied and during occlusion. The bent-mini tip reached temperatures as high as 70°C during occlusion with load when operated at 100% amplitude, whereas the mini-flared tip remained <50°C, and the balanced tip remained <36°C in all test conditions. For the mini-flared and balanced tips, temperature increases during operation were not markedly different from free flow and no-load conditions when occlusion or frictional events were simulated.
Conclusion: In all experiments for each tip design, increasing ultrasound power was associated with greater increases in tip temperature. Tip temperatures increased with applied load, but marked temperature increases during occlusion were observed only with the bent-mini tip. The balanced tip produced minimal thermal peaks in all tests.

Keywords: aspiration bypass, balanced tip, cataract, occlusion, temperature, thermal imaging

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.