Iterative reconstruction for quantitative computed tomography analysis of emphysema: consistent results using different tube currents

Authors Yamashiro T, Miyara T, Honda O, Tomiyama N, Ohno Y, Noma S, Murayama S

Received 23 September 2014

Accepted for publication 1 December 2014

Published 11 February 2015 Volume 2015:10(1) Pages 321—327

DOI https://doi.org/10.2147/COPD.S74810

Checked for plagiarism Yes

Review by Single-blind

Peer reviewer comments 2

Editor who approved publication: Dr Richard Russell


Tsuneo Yamashiro,1 Tetsuhiro Miyara,1 Osamu Honda,2 Noriyuki Tomiyama,2 Yoshiharu Ohno,3 Satoshi Noma,4 Sadayuki Murayama1

On behalf of the ACTIve Study Group

1Department of Radiology, Graduate School of Medical Science, University of the Ryukyus, Nishihara, Okinawa, Japan; 2Department of Radiology, Osaka University Graduate School of Medicine, Suita, Osaka, Japan; 3Department of Radiology, Kobe University Graduate School of Medicine, Kobe, Hyogo, Japan; 4Department of Radiology, Tenri Hospital, Tenri, Nara, Japan

Purpose: To assess the advantages of iterative reconstruction for quantitative computed tomography (CT) analysis of pulmonary emphysema.
Materials and methods: Twenty-two patients with pulmonary emphysema underwent chest CT imaging using identical scanners with three different tube currents: 240, 120, and 60 mA. Scan data were converted to CT images using Adaptive Iterative Dose Reduction using Three Dimensional Processing (AIDR3D) and a conventional filtered-back projection mode. Thus, six scans with and without AIDR3D were generated per patient. All other scanning and reconstruction settings were fixed. The percent low attenuation area (LAA%; < -950 Hounsfield units) and the lung density 15th percentile were automatically measured using a commercial workstation. Comparisons of LAA% and 15th percentile results between scans with and without using AIDR3D were made by Wilcoxon signed-rank tests. Associations between body weight and measurement errors among these scans were evaluated by Spearman rank correlation analysis.
Results: Overall, scan series without AIDR3D had higher LAA% and lower 15th percentile values than those with AIDR3D at each tube current (P<0.0001). For scan series without AIDR3D, lower tube currents resulted in higher LAA% values and lower 15th percentiles. The extent of emphysema was significantly different between each pair among scans when not using AIDR3D (LAA%, P<0.0001; 15th percentile, P<0.01), but was not significantly different between each pair among scans when using AIDR3D. On scans without using AIDR3D, measurement errors between different tube current settings were significantly correlated with patients’ body weights (P<0.05), whereas these errors between scans when using AIDR3D were insignificantly or minimally correlated with body weight.
Conclusion: The extent of emphysema was more consistent across different tube currents when CT scans were converted to CT images using AIDR3D than using a conventional filtered-back projection method.

Keywords: emphysema, iterative reconstruction, Adaptive Iterative Dose Reduction using Three Dimensional Processing, computed tomography



Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.