Evaluation of ultrasonic biomicroscopy results in anterior eye segment before and after cataract surgery

Authors Simsek A, Ciftci S

Received 2 September 2012

Accepted for publication 10 October 2012

Published 21 November 2012 Volume 2012:6 Pages 1931—1934

DOI https://doi.org/10.2147/OPTH.S37614

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 2



Ali Simsek,1 Süleyman Ciftci2

1Department of Ophthalmology, Faculty of Medicine, Adiyaman University, Adiyaman, Turkey; 2Department of Ophthalmology, Diyarbakir Training and Research Hospital, Diyarbakir, Turkey

Background: The aim of this study was to assess the value of ultrasonic biomicroscopy in reporting decreases in intraocular pressure resulting from changes in anterior chamber depth and angle after phacoemulsification and intracapsular lens implantation in patients with cataract.
Methods: This prospective interventional case series included 50 eyes of 50 consecutive subjects operated at the same center. Patients with eye disease affecting visual acuity, a history of eye surgery, corneal surface irregularities, a pupil diameter < 5 mm after preoperative dilation, aged younger than 35 years, posterior capsule perforation, iris dialysis during surgery, intensive postoperative corneal edema, and inability to attend adequate follow-up were excluded. Intraocular pressure, anterior chamber depth and angle, and corneal thickness were measured before and one month after surgery.
Results: The mean preoperative intraocular pressure was 14 mmHg and postoperatively was 11 mmHg. Mean anterior chamber depth preoperatively was 2.8 mm and increased to 3.7 mm postoperatively. The mean anterior chamber angle was measured as 27° preoperatively and as 42° postoperatively.
Conclusion: After phacoemulsification and intracapsular lens implantation, ultrasonic biomicroscopy showed that the iris diaphragm had shifted backwards, widening the angle of the anterior chamber and decreasing intraocular pressure.

Keywords: anterior chamber depth, anterior chamber angle, ultrasonic biomicroscopy

Creative Commons License © 2012 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.