Evaluation of active and passive shimming in magnetic resonance imaging

Authors Wachowicz K

Received 22 April 2014

Accepted for publication 25 June 2014

Published 15 October 2014 Volume 2014:4 Pages 1—12

DOI https://doi.org/10.2147/RRNM.S46526

Checked for plagiarism Yes

Review by Single-blind

Peer reviewer comments 3


Keith Wachowicz

Division of Medical Physics, Department of Oncology, University of Alberta, Edmonton, AB, Canada

Abstract: With higher magnet strengths for magnetic resonance imaging (MRI) units becoming more commonplace, both for animal imaging systems as well as whole-body in vivo systems, magnetic field distortions due to inhomogeneous distributions of magnetic susceptibility and air–tissue interfaces will become more intense. Further, the popularization of MRI-hybrid devices, such as positron emission tomography/MR or MR/radiotherapy hybrids, which rely on the assumption of geometric accuracy for their diagnostic or therapeutic effectiveness, will lead to greater restrictions on permissible geometric error. As a result, shimming procedures (methods by which the distortions induced on the main magnetic field, B0, are remedied) are requiring greater flexibility and corrective range. Shimming methods can be broadly classified into passive (using materials with magnetic properties to remedy field distortions through their passive response to B0) and active techniques (utilizing strategically placed and energized electric coils to produce corrective magnetic fields). Both these techniques have promise to address the additional challenges brought about by the changing MRI landscape. This work reviews traditional shimming methods and principles, and gives an overview of new and novel approaches to this ever-important issue in MRI.

Keywords:
MRI, shim, passive, active, high-order

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.

Download Article [PDF]  View Full Text [HTML][Machine readable]