Back to Journals » Clinical Ophthalmology » Volume 7

Development of a novel disposable lid speculum with a drape

Authors Urano T, Kasaoka M, Yamakawa R, Tamai Y, Nakamura S

Received 12 March 2013

Accepted for publication 29 April 2013

Published 5 August 2013 Volume 2013:7 Pages 1575—1580

DOI https://doi.org/10.2147/OPTH.S45132

Checked for plagiarism Yes

Review by Single-blind

Peer reviewer comments 3

Video abstract presented by Toru Urano.

Views: 3552

Toru Urano,1 Masataka Kasaoka,1 Ryoji Yamakawa,1 YukihikoTamai,2 Shoichiro Nakamura2

1Department of Ophthalmology, Kurume University School of Medicine, Kurume, Japan; 2Hakko Co, Ltd, Nagano, Japan

Purpose: To evaluate the clinical use of a newly-developed disposable lid speculum with a drape.
Methods: LiDrape® is a cylindrical device that consists of two flexible rings of polyacetal resin with a transparent elastic silicone sheet attached to the rings. The novel device holds the eyelids between the rings, and a hole in the center of the device provides a surgical field. We used the novel device in cataract surgery (75 eyes), glaucoma surgery (eleven eyes), vitrectomy (ten eyes), and intravitreal injection (six eyes) and evaluated its clinical efficacy.
Results: The LiDrape was easy to attach and detach. The novel device was not detached from the eye during surgery. No eyelashes or secretions from the meibomian glands were seen in the surgical field, and the drape provided a sufficient surgical field.
Conclusions: The LiDrape functions as a lid speculum as well as a drape. Our results showed that the novel device is useful for ocular surgeries.

Keywords: LiDrape, lid speculum, drape, endophthalmitis

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.

Download Article [PDF]  View Full Text [HTML][Machine readable]

 

Other article by this author:

Evaluation of lid speculum with a drape (LiDrape®) for preventing surgical-field contamination

Urano T, Kasaoka M, Sagawa K, Yamakawa R

Clinical Ophthalmology 2015, 9:1227-1232

Published Date: 8 July 2015

Readers of this article also read:

Green synthesis of water-soluble nontoxic polymeric nanocomposites containing silver nanoparticles

Prozorova GF, Pozdnyakov AS, Kuznetsova NP, Korzhova SA, Emel’yanov AI, Ermakova TG, Fadeeva TV, Sosedova LM

International Journal of Nanomedicine 2014, 9:1883-1889

Published Date: 16 April 2014

Methacrylic-based nanogels for the pH-sensitive delivery of 5-Fluorouracil in the colon

Ashwanikumar N, Kumar NA, Nair SA, Kumar GS

International Journal of Nanomedicine 2012, 7:5769-5779

Published Date: 15 November 2012

A novel preparation method for silicone oil nanoemulsions and its application for coating hair with silicone

Hu Z, Liao M, Chen Y, Cai Y, Meng L, Liu Y, Lv N, Liu Z, Yuan W

International Journal of Nanomedicine 2012, 7:5719-5724

Published Date: 12 November 2012

Cross-linked acrylic hydrogel for the controlled delivery of hydrophobic drugs in cancer therapy

Deepa G, Thulasidasan AK, Anto RJ, Pillai JJ, Kumar GS

International Journal of Nanomedicine 2012, 7:4077-4088

Published Date: 27 July 2012

Crystallization after intravitreal ganciclovir injection

Pitipol Choopong, Nattaporn Tesavibul, Nattawut Rodanant

Clinical Ophthalmology 2010, 4:709-711

Published Date: 14 July 2010