1Chung-Ang University, Seoul, South Korea; 2Augen-Laser-Klinik Nürnberg, Nuremberg, Germany; 3Carl Zeiss Meditec AG, Jena, Germany

Purpose: One hundred eyes from 55 adult patients with myopia were retrospectively studied to determine the comparative safety, efficacy, and predictability of aberration smart ablation (ASA) and a new advanced ablation algorithm (Triple-A) using the MEL® 80 excimer laser.
Methods: Fifty myopic eyes with a manifest refraction spherical equivalent (MRSE) between -1.0 diopters (D) and -9.75 D were consecutively treated with photorefractive keratectomy ASA, and 50 myopic eyes with an MRSE between -1.38 D and -11.0 D with photorefractive keratectomy Triple-A. Uncorrected distance visual acuity, MRSE, the absolute value of the cylinder, corrected distance visual acuity, and postoperative complications at 1 month, 3 months, 6 months, and 12 months (1 year) were descriptively analyzed and compared at 1 year.
Results: After 12 months, the MRSE variance was statistically significantly better in patients triaged to receive Triple-A compared with patients receiving ASA (ASA, ±0.7 D; Triple-A, ±0.15 D; P<0.001). Furthermore, no patient in the Triple-A group had any cylinder postoperatively. Patients in the Triple-A treatment arm achieved a superior result. No statistically significant difference in the two treatment arms was noted for the analysis of the mean MRSE at 12 months (P=0.78).
Conclusion: Triple-A was more effective than standard aspherical surgical intervention in a number of treatment outcome parameters (eg, MRSE, astigmatism, efficacy index). The two surgical procedures were equivalent in terms of safety.

Keywords: aberration smart ablation (ASA), manifest refraction spherical equivalent, Triple-A advanced ablation algorithm, uncorrected distance visual acuity, corrected distance visual acuity, excimer laser, PRK, ablation profile

Creative Commons License This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.