Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes

Authors Alfarhan H , Al-Otaibi

Received 14 April 2012

Accepted for publication 29 May 2012

Published 6 July 2012 Volume 2012:6 Pages 1037—1043

DOI https://doi.org/10.2147/OPTH.S32955

Review by Single anonymous peer review

Peer reviewer comments 2



Haya M Al-Farhan, Wafa’a Majed Al-Otaibi
Department of Optometry and Vision Sciences, College of Applied Medicine Science, King Saud University, Riyadh, Kingdom of Saudia Arabia

Purpose: To compare the precision of central corneal thickness (CCT) measurements taken with the handheld ultrasound pachymeter (USP), ultrasound biomicroscopy (UBM), and the Artemis-2 very high frequency ultrasound scanner (VHFUS) on normal subjects.
Design: Prospective study.
Methods: One eye from each of 61 normal subjects was randomly selected for this study. The measurements of the CCT were taken with the USP, VHFUS, and UBM. Results were compared statistically using repeated-measures analysis of variance (ANOVA), Pearson’s correlation coefficient, and limits of agreement.
Results: The average CCT (± standard deviation) was 530.1 ± 30.5 µm, 554.9 ± 31.7 µm, and 559.5 ± 30.7 µm for UBM, VHFUS, and USP respectively. The intraobserver repeatability analyses of variance are not significant for USP, UBM, and VHFUS. P-values were 0.17, 0.19, and 0.37 respectively. Repeated-measures ANOVA showed a significant difference between the three different methods of measuring CCT (P = 0.0001). The ANOVA test revealed no statistically significant difference between USP and VHFUS (P > 0.05), yet statistical significant differences with UBM versus USP and UBM versus VHFUS (P < 0.001). There were high correlations between the three instruments (P < 0.0001). The mean differences (and upper/lower limits of agreement) for CCT measurements were 29.4 ± 14.3 (2.7/56), 4.6 ± 8.6 (-14.7/23.8), and -24.8 ± 13.1 (-50.4/0.8) for USP versus UBM, USP versus VHFUS, and UBM versus VHFUS, respectively.
Conclusion: The UBM produces CCT measurements that vary significantly from those returned by the USP and the VHFUS, suggesting that the UBM may not be used interchangeably with either equipment for monitoring the CCT in the clinical setting.

Keywords: central corneal thickness, ultrasound pachymetry, ultrasound biomicroscopy, Artemis-2 VHFUS, Artemis, normal eyes

Creative Commons License © 2012 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.