The use of a photoionization detector to detect harmful volatile chemicals by emergency personnel
Original Research
(2230) Views (650) Full article downloads
Authors:
Neil D Patel, William D Fales, Robert N Farrell
Published Date September 2009
Volume 2009:1 Pages 5 - 9
DOI: http://dx.doi.org/10.2147/OAEM.S6855
Neil D Patel1, William D Fales1, Robert N Farrell1,2
1Michigan State University, Kalamazoo Center for Medical Studies, Kalamazoo, MI, USA; 2Portage Fire Department, Portage, MI, USA
Objective: The objective of this investigation was to determine if a photoionization detector (PID) could be used to detect the presence of a simulated harmful chemical on simulated casualties of a chemical release.
Methods: A screening protocol, based on existing radiation screening protocols, was developed for the purposes of the investigation. Three simulated casualties were contaminated with a simulated chemical agent and two groups of emergency responders were involved in the trials. The success–failure ratio of the participants was used to judge the performance of the PID in this application.
Results: A high success rate was observed when the screening protocol was properly adhered to (97.67%). Conversely, the success rate suffered when participants deviated from the protocol (86.31%). With one exception, all failures were noted to have been the result of a failure to correctly observe the established screening protocol.
Conclusions: The results of this investigation indicate that the PID may be an effective screening tool for emergency responders. However, additional study is necessary to both confirm the effectiveness of the PID and refine the screening protocol if necessary.
Keywords: prehospital, device, protocol, photoionization detectors
Readers of this article also read:
Exacerbation rate, health status and mortality in COPD – a review of potential interventions
Knowledge and attitudes towards cardiopulmonary resuscitation and defibrillation amongst Asian primary health care physicians
Dronedarone in the management of atrial fibrillation
Strategies for the prevention and treatment of sudden cardiac death
Antimicrobial prophylaxis in open lower extremity fractures
Solid self-nanoemulsifying cyclosporin A pellets prepared by fluid-bed coating: preparation, characterization and in vitro redispersibility
Emerging treatment options for spasticity in multiple sclerosis – clinical utility of cannabinoids
A review of the evidence concerning hepatic glutathione depletion and susceptibility to hepatotoxicity after paracetamol overdose
Update on pediatric dystonias: etiology, epidemiology, and management
- Testimonials
"... I was impressed at the rapidity of publication from submission to final acceptance." Dr Edwin Thrower, PhD, Yale University
- An eight-week yoga intervention is associated with improvements in pain, psychological functioning and mindfulness, and changes in cortisol levels in women with fibromyalgia
- Problems and barriers of pain management in the emergency department: Are we ever going to get better?
- A new transmucosal drug delivery system for patients with breakthrough cancer pain: the fentanyl effervescent buccal tablet
- Anesthesiologists’ perception of patients’ anxiety under regional anesthesia