The first report on intermediate-term outcome of Ex-PRESS® glaucoma filtration device implanted under scleral flap in Japanese patients

Authors Sugiyama T, Shibata M, Kojima S, Ueki M, Ikeda T

Published 29 July 2011 Volume 2011:5 Pages 1063—1066

DOI https://doi.org/10.2147/OPTH.S23129

Review by Single anonymous peer review

Peer reviewer comments 3



Tetsuya Sugiyama, Maho Shibata, Shota Kojima, Mari Ueki, Tsunehiko Ikeda
Department of Ophthalmology, Osaka Medical College, Takatsuki, Osaka, Japan

Purpose: This paper compares the outcomes of the Ex-PRESS® Glaucoma Filtration Device (Alcon, Fort Worth, TX) implant observed in Japanese patients for 1 year with those of patients undergoing trabeculectomy.
Patients and methods: The subjects comprised ten eyes of ten cases with open-angle glaucoma for which filtration surgery using Ex-PRESS (P-50) was performed by one operator from February 2008 and observed for at least 1 year (Ex-PRESS Group), and eleven eyes of eleven cases for which trabeculectomy was performed by the same operator (TE Group). For both groups, mitomycin C was used and a scleral flap was created after a fornix-based incision of the conjunctiva.
Results: Hypotony and choroidal detachment were observed as early postoperative complications during a 1-week period in one-third of the cases in the TE Group, and failing vision in about 45%, while these were seen in fewer cases in the Ex-PRESS Group. No significant difference in intraocular pressure (IOP) was observed during the period, but IOP variations on the day following the surgery were obviously narrower in the Ex-PRESS Group than in the TE Group. Visual acuity was significantly poorer from 1 week to
3 months in the TE Group while it was stable in the Ex-PRESS Group. The Ex-PRESS Group had fewer cases of laser suture lysis and fewer administrations of glaucoma eyedrop, and no cases of progression in the stage of visual field defect.
Conclusion: Filtration surgery using the Ex-PRESS is unlikely to cause early complications in Japanese patients. Similarly to the trabeculectomy, the intermediate-term control of IOP showed favorable results.

Keywords: Ex-PRESS, miniature glaucoma device, complication, efficacy, Japanese

Creative Commons License © 2011 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.