Back to Journals » Medical Devices: Evidence and Research » Volume 6

Syringeless power injector versus dual-syringe power injector: economic evaluation of user performance, the impact on contrast enhanced computed tomography (CECT) workflow exams, and hospital costs

Authors Colombo GL , Bergamo Andreis IA, Di Matteo S, Bruno GM , Mondellini C

Received 18 July 2013

Accepted for publication 21 September 2013

Published 5 November 2013 Volume 2013:6 Pages 169—174

DOI https://doi.org/10.2147/MDER.S51757

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 3



Giorgio L Colombo,1,2 Ivo A Bergamo Andreis,3 Sergio Di Matteo,2 Giacomo M Bruno,2 Claudio Mondellini3

1Department of Drug Sciences, University of Pavia, Italy; 2Studi Analisi Valutazioni Economiche (S.A.V.E.), Milan, Italy; 3Department of Radiology, City Hospital, Legnano, Italy


Objective: The utilization of diagnostic imaging has substantially increased over the past decade in Europe and North America and continues to grow worldwide. The purpose of this study was to develop an economic evaluation of a syringeless power injector (PI) versus a dual-syringe PI for contrast enhanced computed tomography (CECT) in a hospital setting.
Materials and methods: Patients (n=2379) were enrolled at the Legnano Hospital between November 2012 and January 2013. They had been referred to the hospital for a CECT analysis and were randomized into two groups. The first group was examined with a 256-MDCT (MultiDetector Computed Tomography) scanner using a syringeless power injector, while the other group was examined with a 64-MDCT scanner using a dual-syringe. Data on the operators' time required in the patient analysis steps as well as on the quantity of consumable materials used were collected. The radiologic technologists' satisfaction with the use of the PIs was rated on a 10-point scale. A budget impact analysis and sensitivity analysis were performed under the base-case scenario.
Results: A total of 1,040 patients were examined using the syringeless system, and 1,339 with the dual-syringe system; the CECT examination quality was comparable for both PI systems. Equipment preparation time and releasing time per examination for syringeless PIs versus dual-syringe PIs were 100±30 versus 180±30 seconds and 90±30 and 140±20 seconds, respectively. On average, 10±3 mL of contrast media (CM) wastage per examination was observed with the dual-syringe PI and 0±1 mL with the syringeless PI. Technologists had higher satisfaction with the syringeless PI than with the dual-syringe system (8.8 versus 8.0). The syringeless PI allows a saving of about €6.18 per patient, both due to the lower cost of the devices and to the better performance of the syringeless system. The univariate sensitivity analysis carried out on the base-case results within the standard deviation range confirmed the saving generated by using the syringeless device, with saving values between €5.40 and €6.20 per patient.
Conclusion: The syringeless PI was found to be more user-friendly and efficient, minimizing contrast wastage and providing similar contrast enhancement quality compared to the dual-syringe injector, with comparable CECT examination quality.

Keywords: dual-syringe power injector, syringeless power injector, economic evaluation, cost analysis, computed tomography, CT

Creative Commons License © 2013 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.