New stent design for use in small coronary arteries during percutaneous coronary intervention

Review

(2052) Views  (604) Full article downloads

Authors: Juan F Granada, Barbara A Huibregtse, Keith D Dawkins

Published Date October 2010 Volume 2010:3 Pages 57 - 66
DOI: http://dx.doi.org/10.2147/MDER.S13494

Juan F Granada1, Barbara A Huibregtse2, Keith D Dawkins2
1The Jack H Skirball Center for Cardiovascular Research, Cardiovascular Research Foundation, Columbia University Medical Center, New York, NY, USA; 2Boston Scientific Corporation, Natick, MA, USA

Abstract: Patients with diabetes mellitus, of female gender, increased age, and/or with peripheral vascular disease often develop coronary stenoses in small caliber vessels. This review describes treatment of these lesions with the paclitaxel-eluting 2.25 mm TAXUS® Liberté® Atom™ stent. Given the same stent composition, polymer, antirestenotic drug (paclitaxel), and release kinetics as the first-generation 2.25 mm TAXUS® Express® Atom™ stent, the second-generation TAXUS Liberté Atom stent incorporates improved stent design characteristics, including thinner struts (0.0038 versus 0.0052 inches), intended to increase conformability and deliverability. In a porcine noninjured coronary artery model, TAXUS Liberté Atom stent implantation in small vessels demonstrated complete strut tissue coverage compared with the bare metal stent control, suggesting a similar degree of tissue healing between the groups at 30, 90, and 180 days. The prospective, single-armed TAXUS ATLAS Small Vessel trial demonstrated improved instent late loss (0.28 ± 0.45 versus 0.84 ± 0.57 mm, P < 0.001), instent binary restenosis (13.0% versus 38.1%, P < 0.001), and target lesion revascularization (5.8% versus 17.6%, P < 0.001) at nine months with the TAXUS Liberté Atom stent as compared with the bare metal Express stent control, with similar safety measures between the two groups. The TAXUS Liberté Atom also significantly reduced nine-month angiographic rates of both instent late loss (0.28 ± 0.45 versus 0.44 ± 0.61 mm, P = 0.03) and instent binary restenosis (13.0% versus 25.9%, P = 0.02) when compared with the 2.25 mm TAXUS Express Atom control. The observed reduction in target lesion revascularization with the TAXUS Liberté Atom compared with the TAXUS Express Atom at nine months (5.8% versus 13.7%, P = 0.02) was sustained through three years (10.0% versus 22.1%, P = 0.008) with similar, stable safety outcomes between the groups. In conclusion, these data confirm the safety and favorable performance of the TAXUS Liberté Atom stent in the treatment of small coronary vessels.

Keywords: small vessel, paclitaxel, stent






Readers of this article also read:

Role of aliskiren in cardio-renal protection and use in hypertensives with multiple risk factors
Exacerbation rate, health status and mortality in COPD – a review of potential interventions
ABO and rhesus blood group distribution in Kurds
Anesthesiologists’ perception of patients’ anxiety under regional anesthesia
Everolimus-eluting coronary stents
Improvement of adenoviral vector-mediated gene transfer to airway epithelia by folate-modified anionic liposomes
Perception of risk and benefit in patient-centered communication and care
The relationship between deliberate self-harm behavior, body dissatisfaction, and suicide in adolescents: current concepts
Zinc oxide nanoparticles as selective killers of proliferating cells
Cumulative clinical experience from over a decade of use of levofloxacin in community-acquired pneumonia: critical appraisal and role in therapy
  • Testimonials

    "... I was impressed at the rapidity of publication from submission to final acceptance." Dr Edwin Thrower, PhD, Yale University