Mechanical performances of elastomers used in diffusers

Original Research

(1028) Views  (392) Full article downloads

Authors: Guiffant G, Durussel J-J, Flaud P, Vigier J-P, Dupont C, Bourget P, Merckx J

Published Date June 2011 Volume 2011:4 Pages 71 - 76
DOI: http://dx.doi.org/10.2147/MDER.S18633

Gérard Guiffant1, Jean-Jacques Durussel1, Patrice Flaud1, Jean-Pierre Vigier1, Christian Dupont2, Philippe Bourget3, Jacques Merckx1
1University Paris Diderot, Paris, France; 2University Teaching Hospital Cochin, Paris, France; 3University Teaching Hospital Necker-Enfants Malades, Paris, France

Abstract: The use of elastomeric diffusers (EDs) has grown together with the expansion of home care. In these devices, the fill volume of the drug reservoir and the flow rate are preset and cannot be modified. The elastomer, which makes up the reservoir walls, is what makes the infusate flow due to the pressure it exerts. The purpose of this work was to quantify, under standardized experimental conditions and following recommended conditions of use, the mechanical performances of the 2 commonly used elastomers (silicone and polyisoprene) and their impact on infusion flow rate consistency. Results show that they exhibit different mechanical performances which leads to concerns regarding the use of these devices for some intravenous (IV) therapies.

Keywords: elastomeric diffusers, perfusion, drug administration rate, infusion flow, infusion devices




 

Other articles by Professor Gerard Guiffant



Readers of this article also read:

Periosteoplasty for covering gingival recessions: Clinical results
Role of aliskiren in cardio-renal protection and use in hypertensives with multiple risk factors
Ionically cross-linked hyaluronic acid: wetting, lubrication, and viscoelasticity of a modified adhesion barrier gel
Dealing with thrombocytopenia during anticoagulation with heparins for active venous thromboembolism: a play-it-safe practical approach
Proposal for Alzheimer’s diagnosis using molecular buffer and bus network
Effects of a cellulose mask synthesized by a bacterium on facial skin characteristics and user satisfaction
A better effect of cilostazol for reducing in-stent restenosis after femoropopliteal artery stent placement in comparison with ticlopidine
Medical treatment and long-term outcome of chronic atrial fibrillation in the aged with chest distress: a retrospective analysis versus sinus rhythm
Effects of a transmitted light device for pediatric peripheral venipuncture and intravenous cannulation
Flushing ports of totally implantable venous access devices, and impact of the Huber point needle bevel orientation: experimental tests and numerical computation
  • Testimonials

    "... I was impressed at the rapidity of publication from submission to final acceptance." Dr Edwin Thrower, PhD, Yale University