skip to content
Dovepress - Open Access to Scientific and Medical Research
View our mobile site

8852

Influence of biologically inspired nanometer surface roughness on antigen–antibody interactions for immunoassay–biosensor applications

(2106) Views  (508) Full article downloads

Authors: Paul V Tuttle IV, Ann E Rundell, Thomas J Webster

Published Date May 2006 Volume 2006:1(4) Pages 497 - 505
DOI: http://dx.doi.org/10.2147/IJN.S

Paul V Tuttle IV1, Ann E Rundell1, Thomas J Webster1,2,3

1Weldon School of Biomedical Engineering, and 2School of Materials Engineering, Purdue University, West Lafayette, IN 47907, USA; 3Present address: Division of Engineering, Brown University, Providence, RI, USA

Abstract: Current research efforts to improve immunoassay–biosensor functionality have centered on detection through the optimal design of microfluidic chambers, electrical circuitry, optical sensing elements, and so on. To date, little attention has been paid to the immunoassay–biosensor membrane surface on which interactions between antibodies and antigens must occur. For this reason, the objective of the present study was to manipulate the nanometer surface roughness of a model immunoassay–biosensor membrane to determine its role on sensitivity and specificity. It was hypothesized that surface roughness characteristics similar to those used by the body’s own immune system with B-lymphocyte cell membranes would promote antigen-antibody interactions and minimize non-specific binding. To test this hypothesis, polystyrene 96-well plate surfaces were modified to possess similar topographies as those of B-lymphocyte cell membranes. This was accomplished by immobilizing Protein A conjugated gold particles and Protein A conjugated polystyrene particles ranging in sizes from 40 to 860 nm to the bottom of polystyrene wells. Atomic force microscopy results provided evidence of well-dispersed immunoassay–biosensor surfaces for all particles tested with high degrees of biologically inspired nanometer roughness. Testing the functionality of these immunosurfaces using antigenic fluorescent microspheres showed that specific antigen capture increased with greater nanometer surface roughness while nonspecific antigen capture did not correlate with surface roughness. In this manner, results from this study suggest that large degrees of biologically inspired nanometer surface roughness not only increases the amount of immobilized antibodies onto the immunosurface membrane, but it also enhances the functionality of those antibodies for optimal antigen capture, criteria critical for improving immunoassay–biosensor sensitivity and specificity.

Keywords: nanometer surface roughness, antibody-antigen interaction, B-lymphocyte, immunosurface, immunoassay, biosensor








Readers of this article also read:

Optimizing efficacy of amphotericin B through nanomodification
Intracellular degradation of microspheres based on cross-linked dextran hydrogels or amphiphilic block copolymers: A comparative Raman microscopy study
Enhanced endothelial cell density on NiTi surfaces with sub-micron to nanometer roughness
Evidence-based decision-making within the context of globalization: A “Why–What–How” for leaders and managers of health care organizations
Radiolucency below the crown of mandibular horizontal incompletely impacted third molars and acute inflammation in men with diabetes
Exacerbation rate, health status and mortality in COPD – a review of potential interventions
"Globalized public health.” A transdisciplinary comprehensive framework for analyzing contemporary globalization’s influences on the field of public health
Amino acid management of Parkinson’s disease: a case study
Evaluation of in-vitro antibiotic susceptibility of different morphological forms of Borrelia burgdorferi
Long-term treatment of bipolar disorder with a radioelectric asymmetric conveyor