Comparison of esophageal placement of Bravo capsule system under direct endoscopic guidance with conventional placement method

Authors Sofi A, Filipiak, Sodeman T, Ahmad U, Nawras A, Daboul I

Published 15 October 2010 Volume 2010:3 Pages 147—151

DOI https://doi.org/10.2147/CEG.S12814

Review by Single anonymous peer review

Peer reviewer comments 6



Aijaz A Sofi, Charles Filipiak, Thomas Sodeman, Usman Ahmad, Ali Nawras, Isam Daboul

Department of Medicine, Division of Gastroenterology, University of Toledo Medical Center, Toledo, Ohio, USA

Background: Conventional placement of a wireless esophageal pH monitoring device in the esophagus requires initial endoscopy to determine the distance to the gastroesophageal junction. Blind placement of the capsule by the Bravo delivery system is followed by repeat endoscopy to confirm placement. Alternatively, the capsule can be placed under direct vision during endoscopy. Currently there are no published data comparing the efficiency of one method over the other. The objective of this study was to compare the method of Bravo wireless pH device
placement under direct visualization with the conventional method.
Methods: A retrospective study involving 58 patients (29 patients with indirect and 29 patients with direct visualization) who had Bravo capsule placement. The physician endoscopy procedure notes, nurse’s notes, postprocedure notes, recovery notes, and pH monitoring results were reviewed. The safety of the procedures, length of the procedures, and patient tolerability were evaluated.
Results: None of the 58 patients had early detachment of the device and had no immediate procedure-related complications. The overall incidence of complications in both the groups was similar. No failures due to the technique were noted in either group. Average amount of time taken for the procedure was similar in both groups.
Conclusion: The technique of placing a Bravo pH device under direct visualization is as safe and effective as the conventional method. In addition, there is an added advantage of avoiding a second endoscopic intubation in the direct visualization technique.

Keywords: Bravo capsule, technique, esophageal pH monitoring

Creative Commons License © 2010 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.