Comparison of artifact generation with catheter bending using different PS-OCT approaches

Original Research

(3302) Views  (776) Full article downloads

Authors: Kathy Zheng, Christopher Rashidifard, Bin Liu, Mark Brezinski

Published Date May 2009 Volume 2009:2 Pages 49 - 54
DOI: http://dx.doi.org/10.2147/RMI.S4389

Kathy Zheng1, Christopher Rashidifard1, Bin Liu1,2, Mark Brezinski1,2

1Center for Optical Coherence Tomography and Optical Physics, Department of Orthopedic Surgery, Brigham and Women’s Hospital, Boston, MA, USA; 2Harvard Medical School, Boston, MA, USA

Abstract: Optical coherence tomography (OCT) is a promising biomedical imaging tool for detecting pathology at a micron scale. In addition to performing structural imaging, other adjuvant techniques are available with OCT. In particular, polarization-sensitive optical coherence tomography (PS-OCT) can be used, either through a single or dual detector approach, to assess collagen concentration and organization. In this paper, we compare the ability of single and dual detector PS-OCT with respect to catheter bending artifacts. It was found that even severe catheter bending resulted in no significant difference in quantitative birefringence measurement with the single detector approach. However, when the dual detector approach was examined, catheter bending resulted in a highly significant change in backreflection intensity. This paper raises questions as to the appropriateness of dual detector PS-OCT for performing catheter and endoscope based birefringence assessments.

Keywords: Optical coherence tomography, polarization, single detector, endoscopic imaging

 