Anchoring of self-assembled plasmid DNA/ anti-DNA antibody/cationic lipid micelles on bisphosphonate-modified stent for cardiovascular gene delivery

Authors Ma G, Wang Y, Fishbein I, Yu M, Zhang L, Alferiev IS, Yang J, Song C, Levy RJ, Sun H

Received 7 November 2012

Accepted for publication 12 January 2013

Published 10 March 2013 Volume 2013:8(1) Pages 1029—1035

DOI https://doi.org/10.2147/IJN.S40077

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 2



Guilei Ma,1,# Yong Wang,1,# Ilia Fishbein,2 Mei Yu,1 Linhua Zhang,1 Ivan S Alferiev,2 Jing Yang,1 Cunxian Song,1 Robert J Levy2

1Institute of Biomedical Engineering, Chinese Academy of Medical Sciences and Peking Union Medical College, Tianjin, People's Republic of China; 2Children's Hospital of Philadelphia, Abramson Research Building, Philadelphia, PA, USA

#
These authors contributed equally to this work

Purpose: To investigate the anchoring of plasmid DNA/anti-DNA antibody/cationic lipid tri-complex (DAC micelles) onto bisphosphonate-modified 316 L coronary stents for cardiovascular site-specific gene delivery.
Methods: Stents were first modified with polyallylamine bisphosphonate (PAA-BP), thereby enabling the retention of a PAA-BP molecular monolayer that permits the anchoring (via vector-binding molecules) of DAC micelles. DAC micelles were then chemically linked onto the PAA-BP-modified stents by using N-succinimidyl-3-(2-pyridyldithiol)-propionate (SPDP) as a crosslinker. Rhodamine-labeled DNA was used to assess the anchoring of DAC micelles, and radioactive-labeled antibody was used to evaluate binding capacity and stability. DAC micelles (encoding green fluorescent protein) were tethered onto the PAA-BP-modified stents, which were assessed in cell culture. The presence of a PAA-BP molecular monolayer on the steel surface was confirmed by X-ray photoelectron spectroscopy and atomic force microscope analysis.
Results: The anchoring of DAC micelles was generally uniform and devoid of large-scale patches of defects. Isotopic quantification confirmed that the amount of antibody chemically linked on the stents was 17-fold higher than that of the physical adsorbed control stents and its retention time was also significantly longer. In cell culture, numerous green fluorescent protein-positive cells were found on the PAA-BP modified stents, which demonstrated high localization and efficiency of gene delivery.
Conclusion: The DAC micelle-immobilized PAA-BP-modified stents were successful as a gene delivery system. Gene delivery using DAC micelle-tethered stent-based PAA-BP functionalization should be suitable for a wide array of single or multiple therapeutic gene strategies, and could be used on cardiovascular metallic implants for achieving efficient gene therapy.

Keywords: stent, polyallylamine bisphosphonate, plasmid DNA, micelles, gene delivery

Creative Commons License © 2013 The Author(s). This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution - Non Commercial (unported, v3.0) License. By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms.