An ultra-high input impedance ECG amplifier for long-term monitoring of athletes

Original Research

(3840) Views  (2226) Full article downloads

Authors: Gaetano Gargiulo, Paolo Bifulco, Mario Cesarelli, et al

Published Date July 2010 Volume 2010:3 Pages 1 - 9
DOI: http://dx.doi.org/10.2147/MDER.S9321

Gaetano Gargiulo1,2, Paolo Bifulco1, Mario Cesarelli1, Mariano Ruffo1, Maria Romano1, Rafael A Calvo2, Craig Jin2, André van Schaik2

1Dipartimento di Ingegneria Elettronica, Biomedica e delle Telecomunicazioni, Federico II University of Naples, Italy; 2School of Electrical and Information Engineering, University of Sydney, Australia

Abstract: We present a new, low-power electrocardiogram (ECG) recording system with an ultra-high input impedance that enables the use of long-lasting, dry electrodes. The system incorporates a low-power Bluetooth module for wireless connectivity and is designed to be ­suitable for long-term monitoring during daily activities. The new system using dry electrodes was compared with a clinically approved ECG reference system using gelled Ag/AgCl electrodes and performance was found to be equivalent. In addition, the system was used to monitor an athlete during several physical tasks, and a good quality ECG was obtained in all cases, ­including when the athlete was totally submerged in fresh water.

Keywords: bioelectric potentials, biomedical electrodes, ECG monitoring






Readers of this article also read:

Herpes zoster in the T1 dermatome presenting with Horner’s syndrome, radicular weakness, and postherpetic neuralgia
Berberine: metabolic and cardiovascular effects in preclinical and clinical trials
Potential renovascular hypertension, space missions, and the role of magnesium
Critical appraisal of the role of glucosamine and chondroitin in the management of osteoarthritis of the knee
Ego mechanisms of defense are associated with patients’ preference of treatment modality independent of psychological distress in end-stage renal disease
Can a gentamicin-specific chart reduce neonatal medication errors?
Nephroprotective action of glycosaminoglycans: why the pharmacological properties of sulodexide might be reconsidered
Dashboards in neonatology
Metabolic pathway and distribution of superparamagnetic iron oxide nanoparticles: in vivo study
A quantitative way to estimate clinical off-target effects for human membrane brain targets in CNS research and development
  • Testimonials

    "... I was impressed at the rapidity of publication from submission to final acceptance." Dr Edwin Thrower, PhD, Yale University